CORIO CP-1001F Refrigerated / heating circulator
with an enhanced pump performance and a temperature range up to +200 °C
Refrigerated circulators from the CORIO CP range are suitable for applications with a temperature range up to +200°C. The enhanced pump performance ensures they are suitable for easy temperature control tasks in combination with external applications.
Order No. | 9013708.04 | |||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||||||||||
Heating capacity | 1.8 kW | |||||||||||||||||||||||
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
||||||||||||||||||||||||
Pump capacity flow pressure | 8 ... 27 l/min | |||||||||||||||||||||||
Circulation capacity pressure | 0.1 ... 0.7 bar | |||||||||||||||||||||||
Usable bath opening (W x L / D) |
35 x 41 / 30 cm | |||||||||||||||||||||||
Filling volume | 48 ... 56 l | |||||||||||||||||||||||
Dimensions
(W × L × H) |
45 x 64 x 95 cm |
Product Features
- Models for internal and external applications
- Bright, white, easy to read display
- Very quiet
- powerful and infinitely adjustable pressure pump
- USB connection
- RS232 interface for online communication
- Bath lid and drain tap included
- removable ventilation grid
- Refrigeration unit without side vents
- Class III (FL) according to DIN 12876-1
- easy switching between internal and external circulation
- Integrated pump connection M16×1
- Space-saving cooling coil design yields more usable space in the bath tank
- early warning system for low liquid level
Technical data for CORIO CP-1001F
Dimensions (W × L × H) |
45 x 64 x 95 cm |
---|---|
Filling volume | 48 ... 56 l |
Weight | 73.7 kg |
Pump connections | M16x1 male |
Barbed fittings inner diameter | 8/12 mm |
Working temperature range | -38 ... +100 °C |
---|---|
Temperature stability | ±0.03 °C |
Setting the resolution of the temperature display | 0.1 °C |
Temperature display resolution | 0.01 ... 0.1 °C |
Ambient temperature | +5 ... +40 °C |
Heating capacity | 1.8 kW | |||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
||||||||||||||||||||||||
Pump capacity flow pressure | 8 ... 27 l/min | |||||||||||||||||||||||
Circulation capacity pressure | 0.1 ... 0.7 bar | |||||||||||||||||||||||
Viscosity max. | 50 cSt | |||||||||||||||||||||||
Power consumption | 13 A |
Bath tank | Stainless steel |
---|---|
Bath cover | integrated |
Usable bath opening (W x L / D) |
35 x 41 / 30 cm |
Cooling of compressor | 1-stage Air |
---|
Classification according to DIN12876-1 |
Classification III (FL) |
---|---|
IP Code according to IEC 60529 |
IP 21 |
Pump function | Pressure Pump |
Pump type | Immersion Pump |
Interfaces | RS232 / USB |
---|---|
Temperature control | PID1 |
Absolute temperature calibration | 3 Point Calibration |
Temperature display | LED |
Temperature setting | Keypad |
Electronic Timer h:min |
0 ... 999 |
9013708.04 | 230V/50-60Hz (UK Plug Type BS1363A) (R449A) |
---|---|
9013708.05 | 230V/50-60Hz (CH Plug Type SEV 1011) (R449A) |
9013708.33 | 230V/50-60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A) |
9013708.33.chn | 230V/50-60Hz (CN Plug) (R449A) |
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
Accessory for CORIO CP-1001F
Downloads for CORIO CP-1001F
We would also be happy to send you our printed catalogs and brochures by post.
Request brochureCORIO CP-1001F Benefits
Brilliant.
Very bright display makes it easy to read even from a distance.
Switch on. And off you go.
Intelligent operating concept. Ready for operation with just a few quick and easy steps.
Timer. Integrated.
CORIO circulators include an integrated timer function. When the set time has elapsed, a signal sounds and the device switches off. Setting range: 0 ... 999 minutes.
Everything at the front.
All operating controls and safety functions are accessed easily and comfortably from the front.
Related Products