CORIO CD-900F Refrigerated / heating circulator

The new generation of laboratory circulators for routine work and standard tasks

Refrigerated / heating circulators of the new CORIO series distinguish themselves with a great price-to-performance ratio. They are ideal for all standard tasks and routine work in laboratories and industry.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CD-900F
Order No. 9012706.04
Working temperature range -40 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Cooling capacity
Medium: Ethanol
°C 20 10 0 -10 -20 -30 -40
kW* 0.9
0.9
0.8
0.55
0.35
0.15
0.02
Refrigerant R449A
Global Warming Potential 1397
Filling volume 220 g
Carbon dioxide equivalent 0.307 t
* Performance specifications measured in accordance with DIN12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Usable bath opening
(W x L / D)
26 x 35 / 20 cm
Filling volume 21 ... 30 l
Dimensions
(W × L × H)
39 x 62 x 75 cm

Product Features

  • Models for internal and external applications
  • Bright, white, easy to read display
  • Very quiet
  • Easy pump change-over between internal and external circulation
  • External pump connections
  • USB connection
  • Space-saving cooling coil design yields more usable space in the bath tank
  • Bath lid and drain tap included
  • Removable ventilation grid
  • Refrigeration unit without side vents
  • Class III (FL) according to DIN 12876-1
  • High-quality bath tanks made of stainless steel with bath lid and drain tap
  • Very quiet
Dimensions and volumes
Dimensions
(W × L × H)
39 x 62 x 75 cm
Filling volume 21 ... 30 l
Weight 52 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -40 ... +150 °C
Temperature stability ±0.03 °C
Setting the resolution of the temperature display 0.1 °C
Temperature display resolution 0.01 ... 0.1 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 2 kW
Cooling capacity
Medium: Ethanol
°C 20 10 0 -10 -20 -30 -40
kW* 0.9
0.9
0.8
0.55
0.35
0.15
0.02
Refrigerant R449A
Global Warming Potential 1397
Filling volume 220 g
Carbon dioxide equivalent 0.307 t
* Performance specifications measured in accordance with DIN12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Viscosity max. 50 cST
Power 13 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
26 x 35 / 20 cm
Cooling
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Digital interface USB
Temperature control PID1
Absolute temperature calibration 1 Point Calibration
Temperature display LED
Temperature setting Keypad
Available voltage versions
9012706.02 115V/60Hz (Nema N5-20 Plug) (R449A)
9012706.03 230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A)
9012706.03.chn 230V/50Hz (CN Plug) (R449A)
9012706.04 230V/50Hz (UK Plug Type BS1363A) (R449A)
9012706.05 230V/50Hz (CH Plug Type SEV 1011) (R449A)
9012706.13 230V/60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A)
9012706.12 200V/50-60Hz (Nema N6-20 Plug) (R449A)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Cool-down time

Medium: Ethanol

Heat-up time

Medium: Thermal

Accessory for CORIO CD-900F

We would also be happy to send you our printed catalogs and brochures by post.



Benefits

  • 100% geprüft
    100% Checked.
    100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.
  • ATC.ai
    ATC.
    Absolute Temperature Calibration, 1-point calibration (CD).
  • easy-to-handle
    Handle with ease.
    Makes day-to-day work easy. Comfortably move your JULABO Circulator around by using the ergonomic handles (front and rear).
  • more-bath
    More bath.
    Designed for more comfort. Thanks to the recessed cooling coil, the internal bath provides more space.