Case Study: PRESTO A40 Process system
Heating up of an Normag 10 l reactor from 0 °C to +100 °C with a PRESTO A40 Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO A40 Process system with a 10 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 m metal hoses. The PRESTO A40 Process system has been programmed to heat up from 0 °C to +100 °C.
![Process system PRESTO A40 with QVF glass reactor](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_7b3sf072gh6clcu3a4kk1nbg2g/-S421-FJPG/casestudy-presto-a40-qvf-6-liters-reactor.jpg)
Test Conditions
Product | PRESTO A40 Process system | ||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||
Heating capacity | 2.7 kW | ||||||||||||||
Pump pressure | 0.4 bar | ||||||||||||||
Duration | 1:15 hr:min | ||||||||||||||
Test volume | 10 l | ||||||||||||||
Setpoint 1 | 0 °C | ||||||||||||||
Setpoint 2 | +100 °C | ||||||||||||||
Reactor | 10l Normag (glass) | ||||||||||||||
Tubing length | 2 m | ||||||||||||||
Fluid in unit | Thermal HL40 | ||||||||||||||
Fluid in reactor | Thermal HL40 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 230V/50Hz |