Case Study: PRESTO A45 Process system
Heating up of an Chemglas 20 l reactor from +20 °C to +100 °C with a PRESTO A45 Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO A45 Process system with a 20 l glass reactor. The PRESTO A45 is connected to the reactor via two 2 m metal hoses. The PRESTO A45 Process system has been programmed to heat up from +20 °C to +100 °C.
![Case study Presto A45 10-liter reactor](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_ht2g5ias2d2gd72istj75rhr5v/-S421-FJPG/casestudy-presto-a45-chemglas-10-liters-reactor.jpg)
Test Conditions
Product | PRESTO A45 Process system | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
Heating capacity | 6 kW | ||||||||||||||||
Pump pressure | 0.5 bar | ||||||||||||||||
Duration | 51 hr:min | ||||||||||||||||
Test volume | 15 l | ||||||||||||||||
Setpoint 1 | +20 °C | ||||||||||||||||
Setpoint 2 | +100 °C | ||||||||||||||||
Reactor | 20l Chemglas (glass) | ||||||||||||||||
Tubing length | 2 m | ||||||||||||||||
Fluid in unit | Thermal HL60 | ||||||||||||||||
Fluid in reactor | Thermal H20S |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 400V/3PNPE/50Hz |
Control Parameters
XP | 0.19 K |
TN | 392 s |
TV | 37 s |
XPU | 20 K |
Test Results
The PRESTO A45 Process system heats up the reactor from +20 °C to +100 °C in 51 min.
Set point |
Temperature in the application |
Temperature in temperature control unit |