Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
CORIO CP-1000F
Order number including voltage version
9013707.04
Working temperature range
-50 ... 200 °C
Temperature stability
±0.03 °C
Heating capacity
1.8 kW
Pump capacity flow rate
8 ... 27 l/min
Pump capacity pressure
0.1 ... 0.7 bar
Usable bath opening(W x L / D)
18 x 13 / 15 cm
Filling volume
5 ... 7.5 l
Total dimensions
(W × L × H)
42 x 49 x 74 cm
Models for internal and external applications
Bright, white, easy to read display
Very quiet
powerful and infinitely adjustable pressure pump
USB connection
RS232 interface for online communication
Bath lid and drain tap included
removable ventilation grid
Refrigeration unit without side vents
Class III (FL) according to DIN 12876-1
easy switching between internal and external circulation
Integrated pump connection M16×1
Space-saving cooling coil design yields more usable space in the bath tank
early warning system for low liquid level
Technical data for CORIO CP-1000F
Dimensions and volumes
Total dimensions(W × L × H)
42 x 49 x 74 cm
Filling volume
5 ... 7.5 l
Weight
51.5 kg
Pump connections
M16x1 male
Barbed fittings inner diameter
8/12 mm
Temperature values
Working temperature range
-50 ... 200 °C
Temperature stability
±0.03 °C
Temperature display resolution
0.01 °C
Ambient temperature
+5 ... +40 °C
Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity
1.8 kW
Cooling capacity
°C
20
10
0
-10
-20
-30
-40
kW
1
1
1
0.96
0.73
0.51
0.25
0.11
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate
8 ... 27 l/min
Pump capacity pressure
0.1 ... 0.7 bar
Viscosity max.
50 cSt
Power consumption
13 A
Bath
Bath tank
Stainless steel
Bath cover
integrated
Usable bath opening(W x L / D)
18 x 13 / 15 cm
Cooling
Cooling of compressor
1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function
Pressure Pump
Pump type
Immersion Pump
Electronics
Interfaces
RS232 / USB
Temperature control
PID1
Absolute temperature calibration
3 Point Calibration
Temperature display
LED
Temperature setting
Keypad
Electronic Timer
h:min
0 ... 999
Available voltage versions
9013707.02
115V 60Hz (Nema N5-20 Plug) (R449A)
9013707.04
200-230V 50/60Hz (UK Plug Type BS1363A) (R449A)
9013707.05
200-230V 50/60Hz (CH Plug Type SEV 1011) (R449A)
9013707.33
200-230V 50/60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A)
9013707.33.chn
200-230V 50/60Hz (CN Plug) (R449A)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)
Add to quotation request
Contact expert
Heat-up time
Medium: Thermal
Cool-down time
Medium: Ethanol
Pump capacity
Medium: Water
Accessory for CORIO CP-1000F
Downloads for CORIO CP-1000F
We would also be happy to send you our printed catalogs and brochures by post.
Request brochure