FLW11006 Powerful recirculating cooler

Models with particularly high power reserves

The powerful FL models are suitable for a wide range of cooling tasks in industrial environments, such as removal of large process heat. 2 variants: Air-cooled (FL) and water-cooled (FLW).

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
FLW11006
Order No. 9676110.07
Working temperature range -20 ... +40 °C
Temperature stability ±0.5 °C
Cooling capacity
Medium: Ethanol
°C 20 10 0 -10 -20
kW* 11.5
9
7.3
4.8
2.7
Refrigerant R452A
Global Warming Potential 2140
Filling volume 2330 g
Carbon dioxide equivalent 4.986 t
* Performance specifications measured in accordance with DIN12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate 60 l/min
Pump capacity flow pressure 0.5 ... 6 bar
Filling volume 39 ... 47 l
Dimensions
(W × L × H)
78 x 85 x 148 cm

Product Features

  • Ergonomic design and easy operation
  • Splash-proof keypad
  • Large, bright LED display
  • Permissible temperature in return line +80°C
  • Easy filling from the top with hinged protective lid
  • Low liquid level protection with optical and audible alarm signal
  • Integrated stainless steel bath tanks
  • Front drain
  • No side vents, instruments can be placed right next to other equipment
  • RS232 interface for PC connection
  • IP class according to IEC 60529: 21
  • Alarm output, potential-free change-over contact (max. 30 VA)
  • Pressure Indicator
Dimensions and volumes
Dimensions
(W × L × H)
78 x 85 x 148 cm
Filling volume 39 ... 47 l
Weight 250 kg
Pump connections G1¼" male
Cooling Water Connection G¾ in
Barbed fittings inner diameter 1"
Temperature values
Working temperature range -20 ... +40 °C
Temperature stability ±0.5 °C
Setting the resolution of the temperature display 0.1 °C
Temperature display resolution 0.1 °C
Ambient temperature +5 ... +40 °C
Return flow temperature max. 80 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Cooling capacity
Medium: Ethanol
°C 20 10 0 -10 -20
kW* 11.5
9
7.3
4.8
2.7
Refrigerant R452A
Global Warming Potential 2140
Filling volume 2330 g
Carbon dioxide equivalent 4.986 t
* Performance specifications measured in accordance with DIN12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate 60 l/min
Pump capacity flow pressure 0.5 ... 6 bar
Power 17 A
Bath
Bath tank Stainless steel
Cooling
Cooling of compressor 1-stage Water
Cooling water pressure max. 6 bar
Cooling water difference pressure 2 bar
Cooling water temperature max. 20 °C
Cooling water consumption 18.2 l/min
Other
Sound pressure level 74 dbA
Classification
according to DIN12876-1
Classification I (NFL)
IP Code
according to IEC 60529
IP 21
Pump type Immersion Pump
Electronics
Temperature control PID1
Temperature display LED
Temperature setting Keypad
Available voltage versions
9676110.16 230V/3PPE/60Hz (Without Plug) (R449A)
9676110.07 400V/3PNPE/50Hz (Plug 32A CEE) (R452A)
Included in delivery
2 each barbed fittings for tubing 1" inner dia (pumpconnections G1 1/4" male)

Pump capacity

Accessory for FLW11006

We would also be happy to send you our printed catalogs and brochures by post.