Case Study: PRESTO W91tt Process system
Cooling a Ace Glas 100 l reactor from +20 °C to lowest possible temperature with a PRESTO W91tt Process system
Objective
This case study tests the lowest temperature of the JULABO PRESTO W91tt Process system with a 100 l glass reactor. The PRESTO W91tt is connected to the reactor via two 3 m metal hoses. The PRESTO W91tt Process system cools from +20 °C to -84 °C maximum low temperature.
Test Conditions
| Product | PRESTO W91tt Process system | ||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
| Heating capacity | 36 kW | ||||||||||||||||
| Pump pressure | 0.5 bar | ||||||||||||||||
| Duration | 7:00 hr:min | ||||||||||||||||
| Test volume | 70 l | ||||||||||||||||
| Setpoint 1 | +20 °C | ||||||||||||||||
| Setpoint 2 | -90 °C | ||||||||||||||||
| Reactor | 100l Ace Glas (glass) | ||||||||||||||||
| Tubing length | 3 m | ||||||||||||||||
| Fluid in unit | Thermal P90 | ||||||||||||||||
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 480V/3P(N)PE/60Hz |