CORIO CP-650F Refrigerated / heating circulator new

with an enhanced pump performance and a temperature range up to +200 °C

Refrigerated circulators from the CORIO CP range are suitable for applications with a temperature range up to +200°C. The enhanced pump performance ensures they are suitable for easy temperature control tasks in combination with external applications.

The CORIO CP-650F refrigerated circulator, with a working temperature range of -40 ... 200 °C, delivers a powerful cooling capacity of 0.67 kW at 20 °C despite its very compact design. The cooling machine operates with a natural, environmentally friendly refrigerant. Thanks to the optimized refrigerant cycle, the machine requires only 65 g of refrigerant. This allows for flexible use at the site and shipment via air freight.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CP-650F
Order number including voltage version 9013730.N1.04
Working temperature range -40 ... 200 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Pump capacity flow rate 8 ... 27 l/min
Pump capacity pressure 0.1 ... 0.7 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Total dimensions
(W × L × H)
33 x 47 x 70 cm
Arrange callback

Product Features

  • Models for internal and external applications
  • Bright, white, easy to read display
  • Very quiet
  • powerful and infinitely adjustable pressure pump
  • Bath lid and drain tap included
  • removable ventilation grid
  • Refrigeration unit without side vents
  • easy switching between internal and external circulation
  • early warning system for low liquid level
  • NRTL certification for the U.S. and Canada: Available upon request

Technical data for CORIO CP-650F

Dimensions and volumes
Total dimensions
(W × L × H)
33 x 47 x 70 cm
Filling volume 5 ... 7.5 l
Weight 35.7 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -40 ... 200 °C
Temperature stability ±0.03 °C
Temperature display resolution 0.01 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 2 kW
Cooling capacity
Medium: Ethanol
°C 20 0 -20 -30
kW 1 0.67 0.54 0.26 0.11
Refrigerant R290
Global Warming Potential 3
Filling weight 65 g
Carbon dioxide equivalent 0.000195 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow rate 8 ... 27 l/min
Pump capacity pressure 0.1 ... 0.7 bar
Viscosity max. 50 cSt
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Cooling
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Interfaces RS232 / USB
Temperature control PID1
Absolute temperature calibration 3 Point Calibration
Temperature display LED
Temperature setting Keypad
Electronic Timer
h:min
0 ... 999
Available voltage versions
9013730.N1.02 115V 60Hz (Nema N5-15 Plug) (R290)
9013730.N1.33 200-230V 50/60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R290)
9013730.N1.04 200-230V 50/60Hz (UK Plug Type BS1363A) (R290)
9013730.N1.05 200-230V 50/60Hz (CH Plug Type T12 / 10A) (R290)
9013730.N1.33.chn 200-230V 50/60Hz (Nema N5-15 Plug) (R290)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Accessory for CORIO CP-650F

Downloads for CORIO CP-650F

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

CORIO CP-650F Benefits

Brilliant.

Very bright display makes it easy to read even from a distance.

Connection. Easy.

Inclined pump connections (M16×1) facilitate the connection of applications. Each unit includes 2 barbed fittings of 8/12 mm diameter each.

Everything at the front.

All operating controls and safety functions are accessed easily and comfortably from the front.

Exact.

You can rely on it. PID1 control and 'Active Cooling Control' make the new CORIO precise and perfect.

Show all 10 benefits