Case Study: PRESTO W91tt Process system
Heating up of an Büchiglas 100 l reactor from 0 °C to +20 °C with a PRESTO W91tt Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO W91tt Process system with a 100 l glass reactor. The PRESTO W91tt is connected to the reactor via two 2 m metal hoses. The PRESTO W91tt Process system has been programmed to heat up from 0 °C to +20 °C.

Test Conditions
Product | PRESTO W91tt Process system | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
Heating capacity | 36 kW | ||||||||||||||||
Pump pressure | 0.45 bar | ||||||||||||||||
Duration | 30 hr:min | ||||||||||||||||
Test volume | 100 l | ||||||||||||||||
Setpoint 1 | 0 °C | ||||||||||||||||
Setpoint 2 | +20 °C | ||||||||||||||||
Reactor | 100l Büchiglas (glass) | ||||||||||||||||
Tubing length | 2 m | ||||||||||||||||
Fluid in unit | Thermal HL80 | ||||||||||||||||
Fluid in reactor | Thermal HL80 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 400V/3PNPE/50Hz |