PRESTO W92x Process system

Highest cooling capacity of 27 kW and heating capacity up to 36 kW

Best cooling capacity and best heating capacity: PRESTO W92 are the most powerful systems offering the most modern temperature control technology. Environmental conditions can be simulated or vacuum chambers can be kept at defined temperatures (space conditions). The W92 systems always provide sufficient power.

Data at mains voltage / frequency:
400V/3PNPE/50Hz (Without Plug)
W92x
Order No. 9 421 923.07.X
Working temperature range -92.0 ... +250.0 °C
Temperature stability ±0.05 ... ±0.2 °C
Heating capacity 18 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
200 100 20 0 -20 -30 -40 -60 -80
31 29 27 20 11 10.5 10.5 8 2
Cooling capacity
(Medium: Ethanol)
°C 200 100 20 0 -20 -30 -40 -60 -80
kW 31 29 27 20 11 10.5 10.5 8 2
Pump capacity flow rate 18 ... 70 l/min
Pump capacity flow pressure 0.8 ... 5.5 bar
Dimensions
(W × L × H)
95 x 127 x 190 cm
W92x
Order No. 9 421 923.17.X
Working temperature range -92.0 ... +250.0 °C
Temperature stability ±0.05 ... ±0.2 °C
Heating capacity 18 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
200 100 20 0 -20 -30 -40 -60 -80
31 29 27 20 11 10.5 10.5 8 2
Cooling capacity
(Medium: Ethanol)
°C 200 100 20 0 -20 -30 -40 -60 -80
kW 31 29 27 20 11 10.5 10.5 8 2
Pump capacity flow rate 18 ... 70 l/min
Pump capacity flow pressure 0.8 ... 5.5 bar
Dimensions
(W × L × H)
95 x 127 x 190 cm
Arrange callback Accessory

Product Features

  • Cooling capacity up to 31 kW
  • Rapid heating and cooling
  • Wide working temperature ranges using one bath fluid
  • Designed to quickly compensate exothermic and endothermic reactions
  • Unsurpassed power and efficiency
  • Ambient temperature range +5 °C to +40 °C
  • Space optimized design creates more room directly next to the units
  • Integrated 5.7” industrial color touchscreen displays all essential information and enables simple fingertip control
  • Extensive warning, protection, and monitoring functions with detailed self-explanatory messages
  • ICC cascade control for extraordinary precision, temperature stability ±0.05 °C … ±0.2 °C
  • Intelligent warning and safety functions
  • Electronic level indication
  • Powerful circulating pumps, select a stage or a specific pressure
  • SD-Card slot
  • Connections for USB, Ethernet, RS232, and Alarm output
  • Optional analog connections, RS485, Profibus DP, Modbus
Dimensions and volumes
Dimensions
(W × L × H)
95 x 127 x 190 cm
Internal usable expansion volume 40 l
Minimal process volume 28 l
Active heat exchanger volume 16 l
Weight 870 kg
Pump connections M38x1.5 male
Cooling Water Connection G¾ in
Temperature values
Working temperature range -92.0 ... +250.0 °C
Temperature stability ±0.05 ... ±0.2 °C
Setting the resolution of the temperature display 0.01 °C
Temperature display resolution 0.01 °C
Ambient temperature +5.0 ... +40.0 °C
Data at mains voltage / frequency:
400V/3PNPE/50Hz (Without Plug)
Performance values
Heating capacity 18 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
200 100 20 0 -20 -30 -40 -60 -80
31 29 27 20 11 10.5 10.5 8 2
Cooling capacity
(Medium: Ethanol)
°C 200 100 20 0 -20 -30 -40 -60 -80
kW 31 29 27 20 11 10.5 10.5 8 2
Pump capacity flow rate 18 ... 70 l/min
Pump capacity flow pressure 0.8 ... 5.5 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R404A
Global Warming Potential for R404A 3922
Filling volume 4500 g
Carbon dioxide equivalent 17.649 t
Refrigerant stage 2
Refrigerant R23
Global Warming Potential for R23 14800
Filling volume 1250 g
Carbon dioxide equivalent 18.5 t
Power 46 A
Performance values
Heating capacity 18 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
200 100 20 0 -20 -30 -40 -60 -80
31 29 27 20 11 10.5 10.5 8 2
Cooling capacity
(Medium: Ethanol)
°C 200 100 20 0 -20 -30 -40 -60 -80
kW 31 29 27 20 11 10.5 10.5 8 2
Pump capacity flow rate 18 ... 70 l/min
Pump capacity flow pressure 0.8 ... 5.5 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R404A
Global Warming Potential for R404A 3922
Filling volume 4500 g
Carbon dioxide equivalent 17.649 t
Refrigerant stage 2
Refrigerant R23
Global Warming Potential for R23 14800
Filling volume 1250 g
Carbon dioxide equivalent 18.5 t
Power 38 A
Cooling
Cooling of compressor 2-stage Water
Cooling water pressure max. 6 ... 5.5 bar
Cooling water difference pressure 0.5 bar
Cooling water temperature max. 30 °C
Cooling water consumption 16 ... 43 l/min
Other
Sound pressure level 74 dbA
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump type Gear Pump (Magnetically coupled)
Electronics
External pt100 sensor connection integrated
2nd external Pt100 sensor connection accessory
Integrated programmer 8x60 steps
Temperature control ICC
Absolute temperature calibration 3 Point Calibration
Temperature display 5.7" TFT Touchscreen
Temperature setting Touchscreen
Available voltage versions
400V/3PNPE/50Hz Without Plug 9421923.07
480V/3P(N)PE/60Hz Without Plug 9421923.17

Heat-up time

Accessory for PRESTO W92x

Contact expert

We would also be happy to send you our printed catalogs and brochures by post.


Request brochure

Benefits

  • TCF
    Touch display. Perfect operation.
    With the touch display, the user always has an overview of all values and functions. The intuitive and multilingual menu structure enables perfect control.
  • TCF
    Convenience for several users
    Administrator level for customizing instrument settings, user levels with limited permissions for fast and safe defined access, password protection, all levels adjustable
  • ACC
    100 % Cooling capacity
    ‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures
  • ICC
    Intelligent temperature control.
    Intelligent cascade control - automatic and self-optimizing adaptation of the PID control parameters with external stability of +/- 0.05 °C.