PRESTO A30 Process system

Compact process system down to -40 °C

The A30, A40 and W40 process system combine high performance and a very compact design with the advantages of the PRESTO series for a working temperature range down to -40 °C.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
A30
Order No. 9420300.03
Working temperature range -30.0 ... +250.0 °C
Temperature stability ±0.01 ... ±0.05 °C
Heating capacity 2.7 kW
Cooling capacity
Medium: Ethanol
°C 20 0 -10 -20 -30
kW* 0.5
0.4
0.28
0.17
0.02
Refrigerant R452A
Global Warming Potential 2140
Filling volume 340 g
Carbon dioxide equivalent 0.728 t
Pump capacity flow rate 25 l/min
Pump capacity flow pressure 0.5 bar
Dimensions
(W × L × H)
25 x 59 x 62 cm
Arrange callback Accessory

Product Features

  • Highest performance with small footprint
  • Space-saving design optimizes space utilization in your lab
  • Rapid heating and cooling
  • For highly precise, external temperature applications
  • Wide working temperature ranges without changing fluids
  • Heating capacity up to 2.7 kW
  • Cooling capacity up to 1.2 kW
  • Temperature stability ±0.01 °C ... ±0.05 °C
  • Pump pressure up to 1.7 bar, max. flow rate 40 l/min
  • Built-in 5.7’’ industrial color touchscreen
  • SD-Card slot
  • RS232
  • USB connection
  • Modbus
  • RS485 (Accessory)
  • Profibus DP (Accessory)
  • Second external Pt100 sensor connection for A40 and W40 (accessory)
  • Alarm output
Dimensions and volumes
Dimensions
(W × L × H)
25 x 59 x 62 cm
Internal usable expansion volume 1.5 l
Minimal process volume 2.4 l
Active heat exchanger volume 1.4 l
Weight 62 kg
Pump connections M24x1.5 male
Temperature values
Working temperature range -30.0 ... +250.0 °C
Temperature stability ±0.01 ... ±0.05 °C
Setting the resolution of the temperature display 0.01 °C
Temperature display resolution 0.01 °C
Ambient temperature +5.0 ... +40.0 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 2.7 kW
Cooling capacity
Medium: Ethanol
°C 20 0 -10 -20 -30
kW* 0.5
0.4
0.28
0.17
0.02
Refrigerant R452A
Global Warming Potential 2140
Filling volume 340 g
Carbon dioxide equivalent 0.728 t
Pump capacity flow rate 25 l/min
Pump capacity flow pressure 0.5 bar
Viscosity max. 50 cST
Power 15 A
Cooling
Cooling of compressor 1-stage Air
Other
Sound pressure level 54 dbA
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump type Centrifugal Pump (Magnetically coupled)
Electronics
External pt100 sensor connection integrated
2nd external Pt100 sensor connection not available
Integrated programmer 8x60 steps
Temperature control ICC
Absolute temperature calibration 3 Point Calibration
Temperature display 5.7" TFT Touchscreen
Temperature setting Touchscreen
Available voltage versions
9420300.03 200-230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R452A)
9420300.04 230V/50Hz (UK Plug Type BS1363A) (R452A)
9420300.05 230V/50Hz (CH Plug Type SEV 1011) (R452A)
9420300.14 208V/60Hz (Nema N6-20 Plug) (R449A)

Pump capacity

Heat-up time

Cool-down time

Accessory for PRESTO A30

Contact expert

We would also be happy to send you our printed catalogs and brochures by post.


Request brochure

Benefits

  • TCF
    Touch display. Perfect operation.
    With the touch display, the user always has an overview of all values and functions. The intuitive and multilingual menu structure enables perfect control.
  • TCF
    Convenience for several users
    Administrator level for customizing instrument settings, user levels with limited permissions for fast and safe defined access, password protection, all levels adjustable
  • ACC
    100 % Cooling capacity
    ‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures
  • ICC
    Intelligent temperature control.
    Intelligent cascade control - automatic and self-optimizing adaptation of the PID control parameters with external stability of +/- 0.05 °C.