Case Study: PRESTO W50t Process system
Heating up of an Büchiglas 50 l reactor from -20 °C to +20 °C with a PRESTO W50t Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO W50t Process system with a 50 l glass reactor. The PRESTO W50t is connected to the reactor via two 2 m metal hoses. The PRESTO W50t Process system has been programmed to heat up from -20 °C to +20 °C.
![Process system PRESTO W50 with Büchi glass reactor](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_18rjs6m6f96af91o28p5iv951k/-S421-FJPG/casestudy-presto-w50-buechiglas-50-liters-reactor.jpg)
Test Conditions
Product | PRESTO W50t Process system | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
Heating capacity | 12 kW | ||||||||||||||||
Pump pressure | 0.5 bar | ||||||||||||||||
Duration | 33 hr:min | ||||||||||||||||
Test volume | 37.5 l | ||||||||||||||||
Setpoint 1 | -20 °C | ||||||||||||||||
Setpoint 2 | +20 °C | ||||||||||||||||
Reactor | 50l Büchiglas (glass) | ||||||||||||||||
Tubing length | 2 m | ||||||||||||||||
Fluid in unit | Thermal HL60 | ||||||||||||||||
Fluid in reactor | Thermal HL60 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 400V/3PNPE/50Hz |
Control Parameters
XP | 0.19 K |
TN | 695 s |
TV | 85 s |
XPU | 15 K |
Test Results
The PRESTO W50t Process system heats up the reactor from -20 °C to +20 °C in 33 min.
Set point |
Temperature in the application |
Temperature in temperature control unit |