Case Study: MAGIO MS-1200F Refrigerated / heating circulator
Cooling a Normag 2 l reactor from +80 °C to +20 °C with a MAGIO MS-1200F Refrigerated / heating circulator
Objective
This case study tests the cooling capacity of the JULABO MAGIO MS-1200F Refrigerated / heating circulator with a 2 l glass reactor. The MAGIO MS-1200F is connected to the reactor via two 1 m long metal hoses. The MAGIO MS-1200F Refrigerated / heating circulator has been programmed to cool down from +80 °C to +20 °C.
Test Conditions
| Product | MAGIO MS-1200F Refrigerated / heating circulator | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||
| Heating capacity | 2 kW | ||||||||||||||
| Pump pressure | 0.5 bar | Band limit | °k | ||||||||||||
| Duration | 38 hr:min | Test volume | 2 l | ||||||||||||
| Setpoint 1 | +80 °C | Setpoint 2 | +20 °C | ||||||||||||
| Reactor | 2l Normag (glass) | Tubing length | 1 m | ||||||||||||
| Fluid in unit | Thermal H10 | Fluid in reactor | Thermal H10 | ||||||||||||
Environment
| Room temperature | 22 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 230V/50Hz |
Control Parameters
| XP | 0.11 K |
| TN | 250 s |
| TV | 15 s |
| XPU | 2.5 K |