Case Study: PRESTO W91tt Process system
Cooling a Ace Glas 100 l reactor from +20 °C to lowest possible temperature with a PRESTO W91tt Process system
Objective
This case study tests the lowest temperature of the JULABO PRESTO W91tt Process system with a 100 l glass reactor. The PRESTO W91tt is connected to the reactor via two 3 m metal hoses. The PRESTO W91tt Process system cools from +20 °C to -84 °C maximum low temperature.
![Process system PRESTO W91t AceGlas reactor](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_uqbgoic8ot10f4efu0ono3h76t/-S421-FJPG/casestudy-presto-w91t-aceglas-100l-reactor.jpg)
Test Conditions
Product | PRESTO W91tt Process system | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
Heating capacity | 36 kW | ||||||||||||||||
Pump pressure | 0.5 bar | ||||||||||||||||
Duration | 7:00 hr:min | ||||||||||||||||
Test volume | 70 l | ||||||||||||||||
Setpoint 1 | +20 °C | ||||||||||||||||
Setpoint 2 | -90 °C | ||||||||||||||||
Reactor | 100l Ace Glas (glass) | ||||||||||||||||
Tubing length | 3 m | ||||||||||||||||
Fluid in unit | Thermal P90 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 480V/3P(N)PE/60Hz |