CORIO CD-BC4 Heating Circulator

for internal and external temperature tasks with insulated stainless steel bath tanks and working temperatures from +20 °C to +150 °C

for internal and external temperature applications with stainless steel bath tanks and working temperatures from +20 °C to +150 °C

Request a quote
Data at mains voltage / frequency:
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F)
CD-BC4
Order No. 9 012 504
Working temperature range +20 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 0.8 kW
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.27 bar
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Filling volume 3 ... 4.5 l
Dimensions
(W × L × H)
23 x 41 x 42 cm
CD-BC4
Order No. 9 012 504
Working temperature range +20 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 0.8 kW
Pump capacity flow rate 16 l/min
Pump capacity flow pressure 0.35 bar
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Filling volume 3 ... 4.5 l
Dimensions
(W × L × H)
23 x 41 x 42 cm
CD-BC4
Order No. 9 012 504
Working temperature range +20 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 1 kW
Pump capacity flow rate 16 l/min
Pump capacity flow pressure 0.33 bar
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Filling volume 3 ... 4.5 l
Dimensions
(W × L × H)
23 x 41 x 42 cm
CD-BC4
Order No. 9 012 504
Working temperature range +20 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Filling volume 3 ... 4.5 l
Dimensions
(W × L × H)
23 x 41 x 42 cm
CD-BC4
Order No. 9 012 504
Working temperature range +20 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Filling volume 3 ... 4.5 l
Dimensions
(W × L × H)
23 x 41 x 42 cm
CD-BC4
Order No. 9 012 504
Working temperature range +20 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Filling volume 3 ... 4.5 l
Dimensions
(W × L × H)
23 x 41 x 42 cm
CD-BC4
Order No. 9 012 504
Working temperature range +20 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Filling volume 3 ... 4.5 l
Dimensions
(W × L × H)
23 x 41 x 42 cm
CD-BC4
Order No. 9 012 504
Working temperature range +20 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Pump capacity flow rate 17 l/min
Pump capacity flow pressure 0.43 bar
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Filling volume 3 ... 4.5 l
Dimensions
(W × L × H)
23 x 41 x 42 cm

Product Features

  • Models for internal and external applications
  • Bright, white, easy to read display
  • Very quiet
  • Easy switching between internal and external circulation
  • USB connection
  • High-quality bath tanks made of stainless steel with bath lid and drain tap
  • External pump connections
  • Integrated cooling coil for counter-cooling
  • Easy pump change-over between internal and external circulation
  • For internal and external applications
  • Precise temperature control
Dimensions and volumes
Dimensions
(W × L × H)
23 x 41 x 42 cm
Filling volume 3 ... 4.5 l
Weight 8.8 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range +20 ... +150 °C
Temperature stability ±0.03 °C
Setting the resolution of the temperature display 0.1 °C
Temperature display resolution 0.01 ... 0.1 °C
Ambient temperature +5 ... +40 °C
Data at mains voltage / frequency:
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F)
Performance values
Heating capacity 0.8 kW
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.27 bar
Viscosity max. 50 cST
Power 9 A
Performance values
Heating capacity 0.8 kW
Pump capacity flow rate 16 l/min
Pump capacity flow pressure 0.35 bar
Viscosity max. 50 cST
Power 9 A
Performance values
Heating capacity 1 kW
Pump capacity flow rate 16 l/min
Pump capacity flow pressure 0.33 bar
Viscosity max. 50 cST
Power 10 A
Performance values
Heating capacity 2 kW
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Viscosity max. 50 cST
Power 10 A
Performance values
Heating capacity 2 kW
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Viscosity max. 50 cST
Power 10 A
Performance values
Heating capacity 2 kW
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Viscosity max. 50 cST
Power 10 A
Performance values
Heating capacity 2 kW
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Viscosity max. 50 cST
Power 10 A
Performance values
Heating capacity 2 kW
Pump capacity flow rate 17 l/min
Pump capacity flow pressure 0.43 bar
Viscosity max. 50 cST
Power 10 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Cooling
Cooling coil integrated
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Temperature control PID1
Absolute temperature calibration 1 Point Calibration
Temperature display LED
Temperature setting Keypad
Available voltage versions
100V/50-60Hz Nema N5-15 Plug 9012504.01
115V/60Hz Nema N5-15 Plug 9012504.02
230V/50Hz Schuko Plug - CEE 7/4 Plug Type F 9012504.03
230V/50Hz CN Plug 9012504.03.chn
230V/50Hz UK Plug Type BS1363A 9012504.04
230V/50Hz CH Plug Type SEV 1011 9012504.05
230V/60Hz Schuko Plug - CEE 7/4 Plug Type F 9012504.13
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Accessory for CORIO CD-BC4

We would also be happy to send you our printed catalogs and brochures by post.


Request brochure

Benefits

  • 100% geprüft
    100% Checked.
    100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.
  • ATC.ai
    ATC.
    Absolute Temperature Calibration, 1-point calibration (CD).
  • easy-to-handle
    Handle with ease.
    Makes day-to-day work easy. Comfortably move your JULABO Circulator around by using the ergonomic handles (front and rear).
  • condensation-protection_CORIO
    Condensation protection.
    Superb design solution. Integrated ventilation directs air over the bath lid and minimizes condensation.