Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
CORIO CD-BC12
Order number including voltage version
9012512.03
Working temperature range
1
20 ... 150 °C
Temperature stability
±0.03 °C
Heating capacity
2 kW
Pump capacity flow rate
15 l/min
Pump capacity pressure
0.35 bar
Usable bath opening(W x L / D)
22 x 15 / 20 cm
Filling volume
8.5 ... 12 l
Total dimensions
(W × L × H)
33 x 49 x 47 cm
1 For applications near or below ambient temperature: use a cooling coil or JULABO immersion cooler.
Models for internal and external applications
Bright, white, easy to read display
Very quiet
easy switching between internal and external circulation
High-quality bath tanks made of stainless steel with bath lid and drain tap
Integrated cooling coil for counter-cooling with 9,5mm outer diameter
For internal and external applications
Easy operation
NRTL-certified for the US and Canada
Technical data for CORIO CD-BC12
Dimensions and volumes
Total dimensions(W × L × H)
33 x 49 x 47 cm
Filling volume
8.5 ... 12 l
Weight
12.2 kg
Pump connections
M16x1 male
Barbed fittings inner diameter
8/12 mm
Temperature values
Working temperature range
1
20 ... 150 °C
Temperature stability
±0.03 °C
Temperature display resolution
0.01 °C
Ambient temperature
+5 ... +40 °C
1 For applications near or below ambient temperature: use a cooling coil or JULABO immersion cooler.
Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity
2 kW
Pump capacity flow rate
15 l/min
Pump capacity pressure
0.35 bar
Viscosity max.
50 cSt
Power consumption
10 A
Bath
Bath tank
Stainless steel
Bath cover
integrated
Usable bath opening(W x L / D)
22 x 15 / 20 cm
Cooling
Cooling coil
integrated
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function
Pressure Pump
Pump type
Immersion Pump
Electronics
Interfaces
USB
Temperature control
PID1
Absolute temperature calibration
1 Point Calibration
Temperature display
LED
Temperature setting
Keypad
Available voltage versions
9012512.01
100V 50/60Hz (Nema N5-15 Plug)
9012512.02
115V 60Hz (Nema N5-15 Plug)
9012512.03
230V 50Hz (Schuko Plug - CEE 7/4 Plug Type F)
9012512.03.chn
230V 50Hz (CN Plug)
9012512.04
230V 50Hz (UK Plug Type BS1363A)
9012512.05
230V 50Hz (CH Plug Type T12 / 10A)
9012512.13
230V 60Hz (Nema N5-15 Plug)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)
Add to quotation request
Contact expert
Heat-up time
Medium: Thermal
Pump capacity
Medium: Water
Accessory for CORIO CD-BC12
Downloads for CORIO CD-BC12
We would also be happy to send you our printed catalogs and brochures by post.
Request brochure