CORIO CD-BT19 Heating circulator with open bath and transparent bath tanks

for internal and external applications at working temperatures from +20 °C to +90 °C

The open heating bath circulators of the new CORIO series feature high-quality transparent bath tanks and pump connections.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CD-BT19
Order number including voltage version 9012319.03
Working temperature range 1 20 ... 90 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Pump capacity flow rate 15 l/min
Pump capacity pressure 0.35 bar
Usable bath opening
(W x L / D)
30 x 35 / 15 cm
Filling volume 14 ... 19 l
Total dimensions
(W × L × H)
38 x 58 x 38 cm

1 For applications near or below ambient temperature: use a cooling coil or JULABO immersion cooler.

Arrange callback

Product Features

  • Bright, white, easy to read display
  • Very quiet
  • easy switching between internal and external circulation
  • For internal and external applications
  • NRTL-certified for the US and Canada
  • Bath fluid: water

Technical data for CORIO CD-BT19

Dimensions and volumes
Total dimensions
(W × L × H)
38 x 58 x 38 cm
Filling volume 14 ... 19 l
Weight 8 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range 1 20 ... 90 °C
Temperature stability ±0.03 °C
Temperature display resolution 0.01 °C
Ambient temperature +5 ... +40 °C

1 For applications near or below ambient temperature: use a cooling coil or JULABO immersion cooler.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 2 kW
Pump capacity flow rate 15 l/min
Pump capacity pressure 0.35 bar
Viscosity max. 50 cSt
Power consumption 10 A
Bath
Bath tank Tritan
Bath cover accessory
Usable bath opening
(W x L / D)
30 x 35 / 15 cm
Cooling
Cooling coil accessory
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Interfaces USB
Temperature control PID1
Absolute temperature calibration 1 Point Calibration
Temperature display LED
Temperature setting Keypad
Available voltage versions
9012319.01 100V 50/60Hz (Nema N5-15 Plug)
9012319.02 115V 60Hz (Nema N5-15 Plug)
9012319.03 230V 50Hz (Schuko Plug - CEE 7/4 Plug Type F)
9012319.03.chn 230V 50Hz (CN Plug)
9012319.04 230V 50Hz (UK Plug Type BS1363A)
9012319.05 230V 50Hz (CH Plug Type T12 / 10A)
9012319.13 230V 60Hz (Nema N5-20 Plug)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)
Contact expert

Heat-up time

Medium: Water

Pump capacity

Medium: Water

Accessory for CORIO CD-BT19

Downloads for CORIO CD-BT19

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

CORIO CD-BT19 Benefits

Brilliant.

Very bright display makes it easy to read even from a distance.

Timer. Integrated.

CORIO circulators include an integrated timer function. When the set time has elapsed, a signal sounds and the device switches off. Setting range: 0 ... 999 minutes.

Connection. Easy.

Inclined pump connections (M16×1) facilitate the connection of applications. Each unit includes 2 barbed fittings of 8/12 mm diameter each.

Everything at the front.

All operating controls and safety functions are accessed easily and comfortably from the front.

Show all 19 benefits