CORIO CP-1001F Refrigerated / Heating Circulator
with an enhanced pump performance and a temperature range up to +200 °C
Refrigerated Circulators from the CORIO CP range are suitable for applications with a temperature range up to +200°C. The enhanced pump performance ensures they are suitable for easy temperature control tasks in combination with external applications.
Request a quoteOrder No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 1.5 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Order No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 1.5 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Order No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 1.8 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Order No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 1.8 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Order No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 1.8 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Order No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 1.8 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Order No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 2 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Order No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 2 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Order No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 1.5 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Order No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 1.5 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Order No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 2 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Order No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 2 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Order No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 1.5 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Order No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 1.5 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Order No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 2 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Order No. | 9 013 708 | |||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||
Heating capacity | 2 kW | |||||||||||||||
Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||||
Cooling capacity
(Medium: Ethanol) |
||||||||||||||||
|
||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||
Filling volume | 42 ... 56 l | |||||||||||||||
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
Product Features
- Models for internal and external applications
- Bright, white, easy to read display
- Very quiet
- Easy pump change-over between internal and external circulation
- External pump connections
- Powerful and infinitely adjustable pressure pump
- USB connection
- RS232 interface for online communication
- Space-saving cooling coil design provides more usable space in the bath tank
- Bath lid and drain tap included
- Removable ventilation grid
- Refrigeration unit without side vents
- Unique early warning system for low liquid level
- Class III (FL) according to DIN 12876-1
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
---|---|
Filling volume | 42 ... 56 l |
Weight | 162.5 lbs |
Pump connections | M16x1 male |
Barbed fittings inner diameter | 8/12 mm |
Working temperature range | -38 ... +100 °C |
---|---|
Temperature stability | ±0.03 °C |
Temperature display resolution | 0.01 ... 0.1 °C |
Ambient temperature | +5 ... +40 °C |
Heating capacity | 1.5 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 13 A |
Heating capacity | 1.5 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 13 A |
Heating capacity | 1.8 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 13 A |
Heating capacity | 1.8 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 13 A |
Heating capacity | 1.8 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 10 A |
Heating capacity | 1.8 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 10 A |
Heating capacity | 2 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 10 A |
Heating capacity | 2 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 10 A |
Heating capacity | 1.5 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 16 A |
Heating capacity | 1.5 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 16 A |
Heating capacity | 2 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 16 A |
Heating capacity | 2 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 16 A |
Heating capacity | 1.5 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 16 A |
Heating capacity | 1.5 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 16 A |
Heating capacity | 2 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 16 A |
Heating capacity | 2 kW | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||||
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||||
|
|||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||
Refrigerant stage 1 |
Refrigerant | R449A | |||||||||||||||
Global Warming Potential for R449A | 1397 | ||||||||||||||||
Filling volume | 170 g | ||||||||||||||||
Carbon dioxide equivalent | 0.237 t | ||||||||||||||||
Power | 16 A |
Bath tank | Stainless steel |
---|---|
Bath cover | integrated |
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. |
Cooling of compressor | 1-stage Air |
---|
Classification according to DIN12876-1 |
Classification III (FL) |
---|---|
Pump function | Pressure Pump |
Pump type | Immersion Pump |
Temperature control | PID1 |
---|---|
Absolute temperature calibration | 3 Point Calibration |
Temperature displayTemperature display | LED |
Temperature settingTemperature setting | Keypad |
Electronic Timer hr:min |
0 ... 999 |
230V/50-60Hz UK Plug Type BS1363A | 9013708.04 |
---|---|
230V/50-60Hz CH Plug Type SEV 1011 | 9013708.05 |
230V/50-60Hz Schuko Plug - CEE 7/4 Plug Type F | 9013708.33 |
230V/50-60Hz CN Plug | 9013708.33.chn |
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
Accessory for CORIO CP-1001F
Benefits
-
ATC.
Absolute Temperature Calibration, 1-point calibration (CD). -
Condensation protection.
Superb design solution. Integrated ventilation directs air over the bath lid and minimizes condensation. -
Handle with ease.
Makes day-to-day work easy. Comfortably move your CORIO around by using the ergonomic handles (front and rear). -
Internal. External.
The pump is controlled via a lever located directly below the display. Easily change between internal and external circulation.