CORIO CD-1000FW Refrigerated / heating circulator

The new generation of laboratory circulators for routine work and standard tasks

Refrigerated / heating circulators of the CORIO CD series distinguish themselves with a great price-to-performance ratio. Durable and dependable, they are suitable for a wide range of standard and routine work in both laboratories and industrial environments.

The CORIO CD-1000FW is a water-cooled model.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CD-1000FW
Working temperature range -40 ... +150 °C
Temperature stability ±0.03 °C
Usable bath opening
(W x L / D)
7.1 x 5.1 / 5.9 in.
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
16.5 x 19.3 x 29.1 in.
Arrange callback

Product Features

  • Models for internal and external applications
  • Bright, white, easy to read display
  • Very quiet
  • USB connection
  • Integrated pump connection M16×1
  • easy switching between internal and external circulation
  • Bath lid and drain tap included
  • Removable ventilation grid
  • Refrigeration unit without side vents
  • Class III (FL) according to DIN 12876-1
  • Space-saving cooling coil design provides more usable space in the bath tank

Technical data for CORIO CD-1000FW

Dimensions and volumes
Dimensions
(W × L × H)
16.5 x 19.3 x 29.1 in.
Filling volume 5 ... 7.5 l
Weight 113.5 lbs
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -40 ... +150 °C
Temperature stability ±0.03 °C
Setting the resolution of the temperature display 0.1 °C
Temperature display resolution 0.01 ... 0.1 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
7.1 x 5.1 / 5.9 in.
Cooling
Cooling of compressor 1-stage Water
Cooling water pressure max. 87 psi

Recommended cooling water properties
to provide full cooling capacity according to catalog values
Cooling water temperature range 30 °C
Cooling water difference pressure 4.4 psi
Cooling water consumption 1.1 l/min
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Interfaces USB
Temperature control PID1
Absolute temperature calibration 1 Point Calibration
Temperature display LED
Temperature setting Keypad
Available voltage versions
9012727.13 230V/60Hz (Nema N6-20 Plug) (R449A)
9012727.03.chn 230V/50Hz (CN Plug) (R449A)
9012727.04 230V/50Hz (UK Plug Type BS1363A) (R449A)
9012727.05 230V/50Hz (CH Plug Type SEV 1011) (R449A)
9012727.02 115V/60Hz (Nema N5-20 Plug) (R449A)
9012727.03 230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A)
9012727.12 200V/50-60Hz (Nema N6-20 Plug) (R449A)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)
Contact expert

Cool-down time

Medium: Ethanol

Heat-up time

Medium: Thermal

Pump capacity

Medium: Water

Accessory for CORIO CD-1000FW

Downloads for CORIO CD-1000FW

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

CORIO CD-1000FW Benefits

Brilliant.

Very bright display makes it easy to read even from a distance.

Switch on. And off you go.

Intelligent operating concept. Ready for operation with just a few quick and easy steps.

Timer. Integrated.

CORIO circulators include an integrated timer function. When the set time has elapsed, a signal sounds and the device switches off. Setting range: 0 ... 999 minutes.

Everything at the front.

All operating controls and safety functions are accessed easily and comfortably from the front.

Show all 16 benefits