PRESTO L56w process system new

Energy-efficient process systems down to -56 °C with natural refrigerant

The water-cooled PRESTO L56w can temperature control applications with high power requirements very quickly and efficiently. It has a very high cooling capacity of 55kW at 20 °C, offering large power reserves for demanding external applications across the entire working temperature range. This is particularly true at low temperatures, such as those required in the automotive sector or for reactor temperature control in the pharmaceutical industry. The PRESTO L56w process system uses natural, environmentally friendly refrigerant and was also developed with a focus on energy efficiency. In many application scenarios, this leads to significant operating cost savings and thus to a faster amortization of the acquisition costs. At the same time, the lower energy consumption makes a positive contribution to climate protection. Magnetically coupled centrifugal or gear pumps allow the user to optimally adjust the pump performance to the application, even over long distances and heights, as well as in pressure-sensitive applications. Operation is user-friendly and intuitive via the large 7" touch display or via integration into comprehensive control systems.

Water-glycol up to +150 °C

Connection of an expansion kit ( accessorie) allows pressurized temperature control with water-glycol up to a working temperature of +150 °C.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
PRESTO L56w
Order number including voltage version 9423056.17
Working temperature range -56 ... 250 °C
Temperature stability ±0.05 °C
Heating capacity 27 kW
Pump capacity flow rate 0 ... 180 l/min  depends on options
Pump capacity pressure 1.5 ... 60.9 psi  depends on options
Total dimensions
(W × L × H)
28 x 41.7 x 70.1 in.

Product Features

  • Large 7-inch touch display with intuitive menu navigation
  • Rapid heating and cooling
  • Wide working temperature ranges without changing fluids
  • Climate-friendly thanks to natural refrigerant
  • Wide range of interfaces as standard (additional interfaces available as option)
  • Highly dynamic ICC temperature control, stability to ±0.05 °C

Callback

Available Product Options

Pump
Rotary pump 4.2 bar, 180 l/min, Gear pump 5.5 bar, 80 l/min
Interfaces
CANOpen, EtherCAT, Profibus DP, Profinet, OPC-UA, Stakei
Additional Options
Emergency stop category 0, Emergency stop category 1, Gas sensor

Technical data for PRESTO L56w

Dimensions and volumes
Total dimensions
(W × L × H)
28 x 41.7 x 70.1 in.
Internal usable expansion volume 22 l
Minimal process volume 11 l
Active heat exchanger volume 10 l
Weight 1276.5 lbs
Pump connections M38x1.5 male
Cooling Water Connection G¾ in.
Temperature values
Working temperature range -56 ... 250 °C
Temperature stability ±0.05 °C
Setting the resolution of the temperature display 0.01 °C
Temperature display resolution 0.01 °C
Ambient temperature 5 ... 40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 27 kW
Cooling capacity
Cooling capacity 1 Medium: Ethanol
°C 20 0 -20 -30 -40
kW 1 55 51 25 15 7.2
Cooling capacity 2 Medium: Thermal HL60
°C 20 0 -20 -30 -40
kW 1 55 46 20 11.5 7
Cooling capacity 3 Medium: Water Glycol 40:60
°C 20 0 -20 -30 -40
kW 1 55 50 19.5 9.4 4.7
Cooling capacity 4 Medium: Water Glycol 40:60
°C 20 0 -20 -30 -40
kW 1 55 50 20 11 5.5
Refrigerant R1270
Global Warming Potential 2
Filling weight 2 1300 g
Carbon dioxide equivalent 0.0026 t
Minimum room volume 5756.3 ft³

1 Cooling capacity 4 measured with gear pump.
Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

2 Please observe the applicable transportation regulations.

Pump capacity flow rate 0 ... 180 l/min  depends on options
Pump capacity pressure 1.5 ... 60.9 psi  depends on options
Viscosity max. 100 | 280 cSt  depends on options
Power consumption 63 A
Cooling
Max. heat dissipated by unit into cooling water 90 kW

Recommended cooling water properties
to provide full cooling capacity according to catalog values
Cooling water temperature range 10 ... 15 °C
Cooling water difference pressure 21.8 ... 87 psi

Permissible cooling water properties
Cooling water temperature range 5 ... 35 °C
Cooling water difference pressure 7.3 ... 87 psi
Other
Sound pressure level 67 ... 69 dB(A)
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 20
Pump type Centrifugal Pump (Magnetically coupled)
Electronics
Interfaces Alarm output optional / Ethernet / Modbus TCP/IP / Profibus optional / Profinet optional / REG/EPROG optional / RS232 / RS485 / Stakei optional / Standby-Input optional / USB
External pt100 sensor connection integrated
Integrated programmer 8x60 steps
Temperature control ICC
Absolute temperature calibration 10 Point Calibration
Temperature display 7" TFT Touchscreen
Available voltage versions
9423056.17 460V 3PPE 60Hz (Without Plug) (R1270)
9423056.07 400V 3PPE 50Hz (Plug 63A CEE) (R1270)
Contact expert

Pump capacity

Centrifugal Pump, Medium: Water (20 °C)
Gear Pump, Medium: Water-Glycol 40/60 (-40 °C )

Accessory for PRESTO L56w

Downloads for PRESTO L56w