PRESTO A85t Process system

Powerhouses among the process system

High cooling capacities enable extremely low temperatures down to -85 °C possible. The high heating capacity, particularly with the A85t and the W85t, provides even more flexibility in the application.

The highly dynamic temperature control systems PRESTO are designed for high-precision temperature control for a wide range of applications such as reactor vessels or material stress tests. Moreover, by using efficient components, the process systems can compensate exothermic and endothermic reactions exceptionally fast. Permanent internal monitoring and self-lubricating pumps ensure a long life-time. In addition, numerous interfaces offer many remote control possibilities across networks or for integration into higher-level control systems.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
A85t
Working temperature range -85 ... 250 °C
Temperature stability ±0.05 ... ±0.1 °C
Total dimensions
(W × L × H)
24 x 42.5 x 49.2 in.
Arrange callback

Product Features

  • Cooling capacity up to 2.8 kW
  • External Pt100 sensor connection
  • Connections for alarm-output, Pt100 external sensor and standby-input
  • Analog connections, RS485, Profibus DP (accessory)
  • Built-in 5.7’’ industrial color touchscreen
  • Second external Pt100 sensor connection (accessory)
  • Alarm output
  • Heating capacity up to 15 kW
  • Temperature stability ±0.05 °C ... ±0.1 °C
  • Pump pressure up to 3 bar, max. flow rate 80 l/min
  • Removable ventilation grid

Technical data for PRESTO A85t

Dimensions and volumes
Total dimensions
(W × L × H)
24 x 42.5 x 49.2 in.
Internal usable expansion volume 7 l
Minimal process volume 9.5 l
Active heat exchanger volume 5 l
Weight 804.7 lbs
Pump connections M30x1.5 male
Temperature values
Working temperature range -85 ... 250 °C
Temperature stability ±0.05 ... ±0.1 °C
Setting the resolution of the temperature display 0.01 °C
Temperature display resolution 0.01 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Cooling
Cooling of compressor 2-stage Air
Other
Sound pressure level 69 dB(A)
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 20
Pump type Centrifugal Pump (Magnetically coupled)
Electronics
Interfaces Alarm output / Ethernet / Modbus TCP/IP / Profibus optional / REG/EPROG optional / RS232 / RS485 optional / SD memory card / Standby-Input optional / USB
External pt100 sensor connection integrated
2nd external Pt100 sensor connection accessory
Integrated programmer 8x60 steps
Temperature control ICC
Absolute temperature calibration 3 Point Calibration
Temperature display 5.7" TFT Touchscreen
Temperature setting Touchscreen
Available voltage versions
9420852.07.T 400V 3PNPE 50Hz (Plug 32A CEE) (R507)
9420852.16.T 208-230V 3PPE 60Hz (Without Plug) (R507)
9420852.06.T 230V 3PPE 50Hz (Plug 63A CEE) (R507)
Contact expert

Cool-down time

Medium: Thermal HL

Heat-up time

Medium: Thermal HL

Pump capacity

Accessory for PRESTO A85t

Downloads for PRESTO A85t

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

PRESTO A85t Benefits

Touch display. Perfect operation.

With the touch display, the user always has an overview of all values and functions. The intuitive and multilingual menu structure enables perfect control.

Convenience for several users

Administrator level for customizing instrument settings, user levels with limited permissions for fast and safe defined access, password protection, all levels adjustable

100 % Cooling capacity

‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures

Intelligent temperature control.

Intelligent cascade control - automatic and self-optimizing adaptation of the PID control parameters with external stability of +/- 0.05 °C.

Show all 21 benefits