Case Study: PRESTO A30 Process system

Cooling a QVF 6 l reactor from +20 °C to -10 °C with a PRESTO A30 Process system

Objective

This case study tests the cooling capacity of the JULABO PRESTO A30 Process system with a 6 l glass reactor. The PRESTO A30 is connected to the reactor via two 2 m long metal hoses. The PRESTO A30 Process system has been programmed to cool down from +20 °C to -10 °C.

Test Conditions

Product PRESTO A30 Process system
Heating capacity 2.7 kW
Pump pressure 0.5 bar Band limit °k
Duration 2:00 hr:min Test volume 5 l
Setpoint 1 +20 °C Setpoint 2 -10 °C
Reactor 6l QVF (glass) Tubing length 2 m
Fluid in unit Thermal HL60 Fluid in reactor Thermal HL60

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 230V/50Hz
PRESTO A30 details

Test Results

The PRESTO A30 Process system cools the reactor from +20 °C to -10 °C in 2:00 h.

  • Set point
  • Temperature in the application
  • Temperature in temperature control unit