Case Study: PRESTO A40 Process system

Cooling a Normag 10 l reactor from +200 °C to +100 °C with a PRESTO A40 Process system

Objective

This case study tests the cooling capacity of the JULABO PRESTO A40 Process system with a 10 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 m long metal hoses. The PRESTO A40 Process system has been programmed to cool down from +200 °C to +100 °C.

Test Conditions

Product PRESTO A40 Process system
Heating capacity 2.7 kW
Pump pressure 0.4 bar Band limit °k
Duration 1:10 hr:min Test volume 10 l
Setpoint 1 +200 °C Setpoint 2 +100 °C
Reactor 10l Normag (glass) Tubing length 2 m
Fluid in unit Thermal HL40 Fluid in reactor Thermal HL40

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 230V/50Hz
PRESTO A40 details

Test Results

The PRESTO A40 Process system cools the reactor from +200 °C to +100 °C in 1:10 h.

  • Set point
  • Temperature in the application
  • Temperature in temperature control unit