F81-HL Ultra-Low Refrigerated / Heating Circulator

low temperatures below -50 °C with bath opening for internal and external applications

The Ultra-Low Refrigerated Circulators of the HighTech Series with HL or SL circulator feature powerful pressure and suction pumps. The in-struments provide the entire variety of functions of the professional HighTech Series of circulators.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
F81-HL
Order number including voltage version 9312681.33
Working temperature range -81 ... +100 °C
Temperature stability ±0.02 °C
Heating capacity 1.3 kW
Pump capacity flow pressure 22 ... 26 l/min
Pump capacity pressure 0.4 ... 0.7 bar
Usable bath opening
(W x L / D)
13 x 15 / 16 cm
Filling volume 5 ... 6.5 l
Total dimensions
(W × L × H)
50 x 58 x 89 cm
Arrange callback

Product Features

  • VFD COMFORT DISPLAY
  • LCD DIALOG DISPLAY backlit for convenient interactive operation
  • Keypad for setpoints, warning/safety values and menu functions
  • ICC (Intelligent Cascade Control), self-optimizing temperature control
  • TCF Temperature Control Features to optimize the control behavior
  • ATC3 3-Point-Calibration
  • Pt100 External sensor connection for measurement and control
  • SMART PUMP, electronically adjustable pump stages
  • Unique early warning system for low liquid level
  • Adjustable high temperature cut-out, visible via display
  • RS232/RS485 interface for online communication
  • Integrated programmer for 6 x 60 program steps
  • Connections for solenoid valve and HSP booster pump
  • Proportional cooling control
  • Active Cooling Control

Technical data for F81-HL

Dimensions and volumes
Total dimensions
(W × L × H)
50 x 58 x 89 cm
Filling volume 5 ... 6.5 l
Weight 88 kg
Pump connections M16x1 male
Temperature values
Working temperature range -81 ... +100 °C
Temperature stability ±0.02 °C
Setting the resolution of the temperature display 0.01 °C
Temperature display resolution 0.01 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 1.3 kW
Cooling capacity
Medium: Ethanol
°C 20 0 -20 -40 -60 -80
kW 1 0.45 0.38 0.36 0.32 0.27 0.07
Refrigerant Information Refrigerant stage 1 Refrigerant stage 2
Refrigerant R404A R23
Global Warming Potential 3922 14800
Filling weight 500 g 120 g
Carbon dioxide equivalent 1.961 t 1.776 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow pressure 22 ... 26 l/min
Pump capacity pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cSt
Power consumption 16 A
Bath
Bath cover integrated
Usable bath opening
(W x L / D)
13 x 15 / 16 cm
Cooling
Cooling of compressor 2-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Electronics
Interfaces Profibus optional
External pt100 sensor connection integrated
Integrated programmer 6x60 steps
Temperature control ICC
Absolute temperature calibration 3 Point Calibration
Temperature display VFD
Temperature setting Keypad
Available voltage versions
9312681.33 230V/50-60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R404A)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Accessory for F81-HL

Downloads for F81-HL

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

F81-HL Benefits

Connection of additional equipment

Stakei connections for solenoid valve, HSP booster pump and HST booster heater

Early warning system for high/low temperature limits

Maximum safety for applications, optical and audible alarm, convertible to automated cut-off function

Clever pump system

Reliable and consistent pump capacity, electronically adjustable pump stages

Control of the external application

External Pt100 sensor connection for precise measurement and control directly in the external application

Show all 17 benefits

JULABO customer survey

We appreciate if you take part and help us to further improve. The survey takes only 3 minutes and is anonymous. Thank you!