CORIO CP-900F Refrigerated / heating circulator

with an enhanced pump performance and a temperature range up to +200 °C

Refrigerated circulators from the CORIO CP range are suitable for applications with a temperature range up to +200°C. The enhanced pump performance ensures they are suitable for easy temperature control tasks in combination with external applications.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CP-900F
Order No. 9013706.33
Working temperature range -38 ... +200 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Cooling capacity
Medium: Ethanol
°C 20 10 0 -10 -20 -30
kW* 0.9
0.85
0.8
0.52
0.31
0.11
Refrigerant R449A
Global Warming Potential 1397
Filling volume 220 g
Carbon dioxide equivalent 0.307 t
* Performance specifications measured in accordance with DIN12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate 8 ... 27 l/min
Pump capacity flow pressure 0.1 ... 0.7 bar
Usable bath opening
(W x L / D)
26 x 35 / 20 cm
Filling volume 21 ... 30 l
Dimensions
(W × L × H)
39 x 62 x 75 cm

Product Features

  • Models for internal and external applications
  • Bright, white, easy to read display
  • Very quiet
  • Easy pump change-over between internal and external circulation
  • External pump connections
  • Powerful and infinitely adjustable pressure pump
  • USB connection
  • RS232 interface for online communication
  • Space-saving cooling coil design yields more usable space in the bath tank
  • Bath lid and drain tap included
  • Removable ventilation grid
  • Refrigeration unit without side vents
  • Unique early warning system for low liquid level
  • Class III (FL) according to DIN 12876-1
Dimensions and volumes
Dimensions
(W × L × H)
39 x 62 x 75 cm
Filling volume 21 ... 30 l
Weight 52 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -38 ... +200 °C
Temperature stability ±0.03 °C
Temperature display resolution 0.01 ... 0.1 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 2 kW
Cooling capacity
Medium: Ethanol
°C 20 10 0 -10 -20 -30
kW* 0.9
0.85
0.8
0.52
0.31
0.11
Refrigerant R449A
Global Warming Potential 1397
Filling volume 220 g
Carbon dioxide equivalent 0.307 t
* Performance specifications measured in accordance with DIN12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate 8 ... 27 l/min
Pump capacity flow pressure 0.1 ... 0.7 bar
Viscosity max. 50 cST
Power 16 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
26 x 35 / 20 cm
Cooling
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Digital interface RS232 / USB
Temperature control PID1
Absolute temperature calibration 3 Point Calibration
Temperature display LED
Temperature setting Keypad
Electronic Timer
hr:min
0 ... 999
Available voltage versions
9013706.02 115V/60Hz (Nema N5-20 Plug) (R449A)
9013706.04 200-230V/50-60Hz (UK Plug Type BS1363A) (R449A)
9013706.05 200-230V/50-60Hz (CH Plug Type SEV 1011) (R449A)
9013706.33 200-230V/50-60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A)
9013706.33.chn 200-230V/50-60Hz (CN Plug) (R449A)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Cool-down time

Medium: Ethanol

Accessory for CORIO CP-900F

We would also be happy to send you our printed catalogs and brochures by post.



Benefits

  • condensation-protection_CORIO
    Condensation protection.
    Superb design solution. Integrated ventilation directs air over the bath lid and minimizes condensation.
  • easy-to-handle
    Handle with ease.
    Makes day-to-day work easy. Comfortably move your JULABO Circulator around by using the ergonomic handles (front and rear).
  • easy-pump-control
    Internal. External.
    The pump is controlled via a lever located directly below the display. Easily change between internal and external circulation.
  • easy-to-move
    Mobile.
    Extra easy handling. Integrated castors for easy repositioning of refrigerated circulators.