CORIO CP-200F Refrigerated / heating circulator

with an enhanced pump performance and a temperature range up to +200 °C

Refrigerated circulators from the CORIO CP range are suitable for applications with a temperature range up to +200°C. The enhanced pump performance ensures they are suitable for easy temperature control tasks in combination with external applications.

With a working temperature range of -20 ... 200 °C, the Refrigerated / heating circulator CORIO CP-200F delivers an outstanding cooling capacity of 0.2 kW at 20°C despite its compact design. The optimized design of the bath tank and cooling coil makes the bath more spacious and user-friendly for the operator. Ventilation slots and connections are located on the front and rear, allowing several units to be placed next to each other in order to save space.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CP-200F
Order number including voltage version 9013701.N1.33
Working temperature range -20 ... 200 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Pump capacity flow rate 8 ... 27 l/min
Pump capacity pressure 0.1 ... 0.7 bar
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Filling volume 3 ... 4 l
Total dimensions
(W × L × H)
23 x 39 x 65 cm
Arrange callback

Product Features

  • Models for internal and external applications
  • Bright, white, easy to read display
  • Very quiet
  • powerful and infinitely adjustable pressure pump
  • USB connection
  • RS232 interface for online communication
  • Bath lid and drain tap included
  • removable ventilation grid
  • Refrigeration unit without side vents
  • Class III (FL) according to DIN 12876-1
  • easy switching between internal and external circulation
  • Integrated pump connection M16×1
  • Space-saving cooling coil design yields more usable space in the bath tank
  • early warning system for low liquid level

Technical data for CORIO CP-200F

Dimensions and volumes
Total dimensions
(W × L × H)
23 x 39 x 65 cm
Filling volume 3 ... 4 l
Weight 26 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -20 ... 200 °C
Temperature stability ±0.03 °C
Temperature display resolution 0.01 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 2 kW
Cooling capacity
Medium: Ethanol
°C 20 10 0 -10 -20
kW 1 0.2 0.17 0.15 0.1 0.02
Refrigerant R290
Global Warming Potential 3
Filling weight 26 g
Carbon dioxide equivalent 7.8E-5 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow rate 8 ... 27 l/min
Pump capacity pressure 0.1 ... 0.7 bar
Viscosity max. 50 cSt
Power consumption 12 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Cooling
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Interfaces RS232 / USB
Temperature control PID1
Absolute temperature calibration 3 Point Calibration
Temperature display LED
Temperature setting Keypad
Electronic Timer
h:min
0 ... 999
Available voltage versions
9013701.N1.33.chn 230V 50/60Hz (CN Plug) (R290)
9013701.N1.33 230V 50/60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R290)
9013701.N1.05 230V 50/60Hz (CH Plug Type T12 / 10A) (R290)
9013701.N1.04 230V 50/60Hz (UK Plug Type BS1363A) (R290)
9013701.N1.02 115V 60Hz (Nema N5-15 Plug) (R290)
9013701.N1.01 100V 50/60Hz (Nema N5-15 Plug) (R290)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)
Contact expert

Heat-up time

Medium: Thermal

Cool-down time

Medium: Ethanol

Pump capacity

Medium: Water

Accessory for CORIO CP-200F

Downloads for CORIO CP-200F

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

CORIO CP-200F Benefits

Brilliant.

Very bright display makes it easy to read even from a distance.

Connection. Easy.

Inclined pump connections (M16×1) facilitate the connection of applications. Each unit includes 2 barbed fittings of 8/12 mm diameter each.

Everything at the front.

All operating controls and safety functions are accessed easily and comfortably from the front.

Exact.

You can rely on it. PID1 control and 'Active Cooling Control' make the new CORIO precise and perfect.

Show all 10 benefits