Case Study: PRESTO W55 Process system
Heating up of an Büchiglas 50 l reactor from +20 °C to +150 °C with a PRESTO W55 Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO W55 Process system with a 50 l glass reactor. The PRESTO W55 is connected to the reactor via two 2 m metal hoses. The PRESTO W55 Process system has been programmed to heat up from +20 °C to +150 °C.
![Process system PRESTO W55 with Büchi glass reactor 50 liters](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_3qg7ukpp793m783jugca0ipb1l/-S421-FJPG/casestudy-presto-w55-buechiglas-50-liters-reactor.jpg)
Test Conditions
Product | PRESTO W55 Process system | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
Heating capacity | 15 kW | ||||||||||||||||
Pump pressure | 0.5 bar | ||||||||||||||||
Duration | 46 hr:min | ||||||||||||||||
Test volume | 35 l | ||||||||||||||||
Setpoint 1 | +20 °C | ||||||||||||||||
Setpoint 2 | +150 °C | ||||||||||||||||
Reactor | 50l Büchiglas (glass) | ||||||||||||||||
Tubing length | 2 m | ||||||||||||||||
Fluid in unit | Thermal HL60 | ||||||||||||||||
Fluid in reactor | Thermal H20S |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 400V/3PNPE/50Hz |
Control Parameters
XP | 0.1 K |
TN | 385 s |
TV | 21 s |
XPU | 12 K |
Test Results
The PRESTO W55 Process system heats up the reactor from +20 °C to +150 °C in 46 min.
Set point |
Temperature in the application |
Temperature in temperature control unit |