Case Study: PRESTO A45t Process system

Heating up of an Asahi 20 l reactor from -20 °C to +20 °C with a PRESTO A45t Process system

Objective

This case study tests the heating capacity of the JULABO PRESTO A45t Process system with a 20 l glass reactor. The PRESTO A45t is connected to the reactor via two 2 m metal hoses. The PRESTO A45t Process system has been programmed to heat up from -20 °C to +20 °C.

Test Conditions

Product PRESTO A45t Process system
Cooling capacity (Ethanol)
°C 200 20 0 -10 -20 -30 -40
kW 3.4 3.5 3.3 2.6 1.8 1 0.3
Heating capacity 12 kW
Pump pressure 0.5 bar Band limit °k
Duration 14 hr:min Test volume 15 l
Setpoint 1 -20 °C Setpoint 2 +20 °C
Reactor 20l Asahi (glass) Tubing length 2 m
Fluid in unit Thermal HL60 Fluid in reactor Thermal HL60

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 400V/3PNPE/50Hz

Control Parameters

XP 0.19 K
TN 600 s
TV 80 s
XPU 15 K
PRESTO A45t details

Test Results

The PRESTO A45t Process system heats up the reactor from -20 °C to +20 °C in 14 min.

  • Set point
  • Temperature in the application
  • Temperature in temperature control unit