Case Study: PRESTO A45t Process system
Cooling a Chemglas 10 l reactor from +20 °C to lowest possible temperature with a PRESTO A45t Process system
Objective
This case study tests the lowest temperature of the JULABO PRESTO A45t Process system with a 10 l glass reactor. The PRESTO A45t is connected to the reactor via two m metal hoses. The PRESTO A45t Process system cools from +20 °C to -39.05 °C maximum low temperature.
![Case study Presto A45 10-liter reactor](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_ht2g5ias2d2gd72istj75rhr5v/-S421-FJPG/casestudy-presto-a45-chemglas-10-liters-reactor.jpg)
Test Conditions
Product | PRESTO A45t Process system | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
Heating capacity | 12 kW | ||||||||||||||||
Pump pressure | 0.5 bar | ||||||||||||||||
Duration | 2:00 hr:min | ||||||||||||||||
Test volume | 7.5 l | ||||||||||||||||
Setpoint 1 | +20 °C | ||||||||||||||||
Setpoint 2 | -55 °C | ||||||||||||||||
Reactor | 10l Chemglas (glass) | ||||||||||||||||
Fluid in unit | Thermal HL60 | ||||||||||||||||
Fluid in reactor | Thermal HL60 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 400V/3PNPE/50Hz |
Control Parameters
XP | 0.19 K |
TN | 300 s |
TV | 40 s |
XPU | 8 K |
Test Results
The PRESTO A45t Process system cools the reactor from +20 °C to the lowest temperature of -39.05 °C in 2:00 h.
Set point |
Temperature in the application |
Temperature in temperature control unit |