11 kW cooling capacity for dynamic temperature changes down to -91 °C
Best heating capacity combined with high cooling capacity – those are the key features of the W91 units. The units are just as ready for embedding into pilot plants as they are for use in material and component testing.
The highly dynamic temperature control systems PRESTO are designed for high-precision temperature control for a wide range of applications such as reactor vessels or material stress tests. Moreover, by using efficient components, the process systems can compensate exothermic and endothermic reactions exceptionally fast. Permanent internal monitoring and self-lubricating pumps ensure a long life-time. In addition, numerous interfaces offer many remote control possibilities across networks or for integration into higher-level control systems.
Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate
0 ... 70 l/min
Pump capacity pressure
1.5 ... 79.8 psi
Viscosity max.
70 cSt
Power consumption
40 A
Cooling
Cooling of compressor
2-stage Water
Cooling water pressure max.
87 psi
Max. heat dissipated by unit into cooling water
28 kW
Recommended cooling water properties
to provide full cooling capacity according to catalog values
With the touch display, the user always has an overview of all values and functions.
The intuitive and multilingual menu structure enables perfect control.
Convenience for several users
Administrator level for customizing instrument settings, user levels with limited permissions for fast and safe defined access, password protection, all levels adjustable
100 % Cooling capacity
‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures
Intelligent temperature control.
Intelligent cascade control - automatic and self-optimizing adaptation of the PID control parameters with external stability of +/- 0.05 °C.
On Friday, 7 August 2026, we will be unavailable by phone and email from 2:30 PM for approximately 45 minutes. We'll be back from around 3:15 PM (CEST).