MAGIO MS-1000F Refrigerated / Heating circulator

with extremely powerful pressure/suction pump and intuitive touch display

As with all circulators from the MAGIO range, the refrigerated circulators stand out thanks to their premium quality, high performance and intuitive operation. The devices offer extra strong pressure and suction pumps, thus fulfilling the highest demands for temperature control of external applications. Whether in basic research, material testing or technical systems – the MAGIO refrigerated circulators offer high-tech solutions for high customer requirements.

Request a quote

High resolution TFT touch display

The modern TFT touch display gives you all the important information at a glance. Three large, predefined main screens clearly display data and graphics with various application priorities. Menu navigation is self-explanatory, arranged by relevance to daily operations and easy to operate with the touch of a finger. The in-built help function provides detailed support in case of additional questions.
Touch_Display
Data at mains voltage / frequency:
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F)
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 1 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 1.6 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 1.6 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 1.6 kW
Cooling capacity
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 1.6 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 1.6 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 1.6 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 1.6 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 1.6 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm
MS-1000F
Order No. 9 032 707
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 70 cm

Product Features

  • Ideal for demanding external applications
  • Simple control of complex applications
  • Continuously adjustable, extremely powerful pressure / suction pump
  • Flow rate 16 ... 31 l / min, pressure 0.24 ... 0.92 bar, suction 0.03 ... 0.4 bar
  • Large, high-resolution TFT touch display with multilingual user interface
  • Parts being in contact with the medium made of stainless steel
  • Integrated programmer
  • Integrated external Pt100 connection
  • USB connection
  • RS232 interface for online communication
  • Ethernet
  • analog interfaces (accessory)
  • Class III (FL) according to DIN 12876-1
  • Modbus
  • Profibus DP (Accessory)
  • RS232/RS485 interface for online communication
  • Connections for solenoid valve
Dimensions and volumes
Dimensions
(W × L × H)
42 x 49 x 70 cm
Filling volume 5 ... 7.5 l
Weight 54.1 kg
Pump connections M16x1 male
Temperature values
Working temperature range -50 ... +200.0 °C
Temperature stability +/-0.01 °C
Setting the resolution of the temperature display 0.01 °C
Temperature display resolution 0.01 °C
Ambient temperature +10.0 ... +40.0 °C
Data at mains voltage / frequency:
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F)
Performance values
Heating capacity 1 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 16 A
Performance values
Heating capacity 1.6 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 10 A
Performance values
Heating capacity 1.6 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 10 A
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 10 A
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 10 A
Performance values
Heating capacity 1.6 kW
Cooling capacity
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 12 A
Performance values
Heating capacity 1.6 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 12 A
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 13 A
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 13 A
Performance values
Heating capacity 1.6 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.04 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 15 A
Performance values
Heating capacity 1.6 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 15 A
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 16 A
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 16 A
Performance values
Heating capacity 1.6 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 13 A
Performance values
Heating capacity 1.6 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 13 A
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 14 A
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -10 -20 -30 -40
1 0.96 0.96 0.7 0.51 0.25 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -10 -20 -30 -40
kW 1 0.96 0.96 0.7 0.51 0.25 0.11
Pump capacity flow rate 16 ... 31 l/min
Pump capacity flow pressure 0.24 ... 0.92 bar
Maximum suction -0.03 ... -0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Power 14 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Cooling
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Suction Pump
Pump type Immersion Pump
Electronics
External pt100 sensor connection integrated
Integrated programmer 8x60 steps
Temperature control ICC
Absolute temperature calibration 3 Point Calibration
Temperature display 7" TFT Touchscreen
Temperature setting Touchscreen
Electronic Timer
hr:min
00:00 ... 99:59
Available voltage versions
115V/60Hz Nema N5-20 Plug 9032707.02
200-230V/50-60Hz CH Plug Type SEV 1011 9032707.05
200-230V/50-60Hz UK Plug Type BS1363A 9032707.04
200-230V/50-60Hz Schuko Plug - CEE 7/4 Plug Type F 9032707.33
200-230V/50-60Hz CN Plug 9032707.33.chn
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Heat-up time

Accessory for MAGIO MS-1000F

We would also be happy to send you our printed catalogs and brochures by post.


Request brochure

Benefits

  • 100% geprüft
    100% Checked.
    100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.
  • Blatt wächst in einer Glühbirne
    Green technology.
    Development consistently applied environmentally friendly materials and technologies.
  • ICC
    Intelligent temperature control.
    Intelligent cascade control - automatic and self-optimizing adaptation of the PID control parameters with external stability of +/- 0.05 °C.
  • Made in Germany
    JULABO. Quality.
    Highest standards of quality for a long product life.