Case Study: PRESTO A40 Process system
Heating up of an QVF 6 l reactor from 0 °C to +100 °C with a PRESTO A40 Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO A40 Process system with a 6 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 m metal hoses. The PRESTO A40 Process system has been programmed to heat up from 0 °C to +100 °C.
Test Conditions
| Product | PRESTO A40 Process system | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||
| Heating capacity | 2.7 kW | ||||||||||||||
| Pump pressure | 0.4 bar | ||||||||||||||
| Duration | 1:15 hr:min | ||||||||||||||
| Test volume | 10 l | ||||||||||||||
| Setpoint 1 | 0 °C | ||||||||||||||
| Setpoint 2 | +100 °C | ||||||||||||||
| Reactor | 6l QVF (glass) | ||||||||||||||
| Tubing length | 2 m | ||||||||||||||
| Fluid in unit | Thermal HL40 | ||||||||||||||
| Fluid in reactor | Thermal HL40 | ||||||||||||||
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 230V/50Hz |