Case Study: PRESTO A80t Process system
Cooling a Asahi 20 l reactor from +20 °C to -20 °C with a PRESTO A80t Process system
Objective
This case study tests the cooling capacity of the JULABO PRESTO A80t Process system with a 20 l glass reactor. The PRESTO A80t is connected to the reactor via two 1 m long metal hoses. The PRESTO A80t Process system has been programmed to cool down from +20 °C to -20 °C.
Test Conditions
| Product | PRESTO A80t Process system | ||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||||||
| Heating capacity | 2.8 kW | ||||||||||||||||||||
| Pump pressure | 0.5 bar | ||||||||||||||||||||
| Duration | 1:25 hr:min | ||||||||||||||||||||
| Test volume | 19 l | ||||||||||||||||||||
| Setpoint 1 | +20 °C | ||||||||||||||||||||
| Setpoint 2 | -20 °C | ||||||||||||||||||||
| Reactor | 20l Asahi (glass) | ||||||||||||||||||||
| Tubing length | 1 m | ||||||||||||||||||||
| Fluid in unit | Thermal HL80 | ||||||||||||||||||||
| Fluid in reactor | Thermal HL80 | ||||||||||||||||||||
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 208V/60Hz |