CORIO CP-1000FW Refrigerated / heating circulator

with an enhanced pump performance and a temperature range up to +200 °C

Refrigerated circulators from the CORIO CP range are suitable for applications with a temperature range up to +200°C. The enhanced pump performance ensures they are suitable for easy temperature control tasks in combination with external applications.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CP-1000FW
Order No. 9013727.33
Working temperature range -50 ... 200 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Cooling capacity
Medium: Ethanol
°C 20 10 0 -10 -20 -30 -40
kW* 1
1
0.96
0.73
0.51
0.25
0.11
Refrigerant R449A
Global Warming Potential 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Pump capacity flow rate 8 ... 27 l/min
Pump capacity flow pressure 0.1 ... 0.7 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
42 x 49 x 74 cm
Arrange callback Accessory

Product Features

  • Models for internal and external applications
  • Bright, white, easy to read display
  • Very quiet
  • Easy pump change-over between internal and external circulation
  • External pump connections
  • Powerful and infinitely adjustable pressure pump
  • USB connection
  • RS232 interface for online communication
  • Space-saving cooling coil design yields more usable space in the bath tank
  • Bath lid and drain tap included
  • Removable ventilation grid
  • Refrigeration unit without side vents
  • Unique early warning system for low liquid level
  • Class III (FL) according to DIN 12876-1
Dimensions and volumes
Dimensions
(W × L × H)
42 x 49 x 74 cm
Filling volume 5 ... 7.5 l
Weight 51.5 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -50 ... 200 °C
Temperature stability ±0.03 °C
Temperature display resolution 0.01 ... 0.1 °C
Ambient temperature 5 ... 40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 2 kW
Cooling capacity
Medium: Ethanol
°C 20 10 0 -10 -20 -30 -40
kW* 1
1
0.96
0.73
0.51
0.25
0.11
Refrigerant R449A
Global Warming Potential 1397
Filling volume 190 g
Carbon dioxide equivalent 0.265 t
Pump capacity flow rate 8 ... 27 l/min
Pump capacity flow pressure 0.1 ... 0.7 bar
Viscosity max. 50 cST
Power 16 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Cooling
Cooling of compressor 1-stage Water
Cooling water pressure max. 6 bar
Cooling water difference pressure 0.3 bar
Cooling water temperature max. 30 °C
Cooling water consumption 1.1 l/min
Other
Classification
according to DIN12876-1
Classification III (FL)
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Digital interface RS232 / USB
Temperature control PID1
Absolute temperature calibration 3 Point Calibration
Temperature display LED
Temperature setting Keypad
Electronic Timer
hr:min
0 ... 999
Available voltage versions
9013727.02 115V/60Hz (Nema N5-20 Plug) (R449A)
9013727.04 200-230V/50-60Hz (UK Plug Type BS1363A) (R449A)
9013727.05 200-230V/50-60Hz (CH Plug Type SEV 1011) (R449A)
9013727.33 200-230V/50-60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A)
9013727.33.chn 200-230V/50-60Hz (CN Plug) (R449A)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Heat-up time

Accessory for CORIO CP-1000FW

Contact expert

We would also be happy to send you our printed catalogs and brochures by post.


Request brochure

Benefits

  • Sicherheit
    Safety.
    CORIO CD and CP comply with Class III (FL) according to DIN 12876-1 and switches off automatically in case of high temperature or low liquid level alarm.
  • brilliant-display
    Brilliant.
    Very bright display makes it easy to read even from a distance.
  • easy-conectivity
    Connection. Easy.
    Inclined pump connections (M16×1) facilitate the connection of applications. Each unit includes 2 barbed fittings of 8/12 mm diameter each.
  • all-in-front
    Everything at the front.
    All operating controls and safety functions are accessed easily and comfortably from the front.