Case Study: PRESTO W92tt Process system
Heating up of an Büchiglas 100 l reactor from -50 °C to +20 °C with a PRESTO W92tt Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO W92tt Process system with a 100 l glass reactor. The PRESTO W92tt is connected to the reactor via two 3 m metal hoses. The PRESTO W92tt Process system has been programmed to heat up from -50 °C to +20 °C.
Test Conditions
| Product | PRESTO W92tt Process system | ||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||||||
| Heating capacity | 36 kW | ||||||||||||||||||||
| Pump pressure | 0.5 bar | ||||||||||||||||||||
| Duration | 1:35 hr:min | ||||||||||||||||||||
| Test volume | 70 l | ||||||||||||||||||||
| Setpoint 1 | -50 °C | ||||||||||||||||||||
| Setpoint 2 | +20 °C | ||||||||||||||||||||
| Reactor | 100l Büchiglas (glass) | ||||||||||||||||||||
| Tubing length | 3 m | ||||||||||||||||||||
| Fluid in unit | Thermal HL80 | ||||||||||||||||||||
| Fluid in reactor | Thermal HL80 | ||||||||||||||||||||
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 400V/3PNPE/50Hz |