Case Study: PRESTO A80 Process system

Heating up of an Asahi 20 l reactor from -20 °C to 0 °C with a PRESTO A80 Process system

Objective

This case study tests the heating capacity of the JULABO PRESTO A80 Process system with a 20 l glass reactor. The PRESTO A80 is connected to the reactor via two 2 m metal hoses. The PRESTO A80 Process system has been programmed to heat up from -20 °C to 0 °C.

Test Conditions

Product PRESTO A80 Process system
Cooling capacity (Ethanol)
°C 20 0 -20 -40 -60 -80
kW 1.2 1.2 1.1 1.1 0.65 0.1
Heating capacity 1.8 kW
Pump pressure 0.4 bar Band limit °k
Duration 36 hr:min Test volume 18 l
Setpoint 1 -20 °C Setpoint 2 0 °C
Reactor 20l Asahi (glass) Tubing length 2 m
Fluid in unit Thermal HL80 Fluid in reactor

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 230V/50Hz
PRESTO A80 details

Test Results

The PRESTO A80 Process system heats up the reactor from -20 °C to 0 °C in 36 min.

  • Set point
  • Temperature in the application
  • Temperature in temperature control unit