CORIO CD-800F Refrigerated / heating circulator new
The new generation of laboratory circulators for routine work and standard tasks
Refrigerated / heating circulators of the CORIO CD series distinguish themselves with a great price-to-performance ratio. Durable and dependable, they are suitable for a wide range of standard and routine work in both laboratories and industrial environments.
With a working temperature range of -40 ... +150 °C, the Refrigerated / heating circulator CORIO CD-800F delivers an outstanding cooling capacity of 0.85 kW at 20 °C despite its compact design.
This cooling machine works with natural, environmentally-friendly refrigerant and was developed with a focus on energy efficiency. This means significant savings on the operating costs for numerous applications, which also means rapid amortization of the procurement cost. At the same time, the lower energy consumption positively contributes to climate protection.
Order No. | 9012715.N1.04 | |||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -40 ... +150 °C | |||||||||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||||||||
Heating capacity | 2 kW | |||||||||||||||||||||
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
||||||||||||||||||||||
Pump capacity flow pressure | 15 l/min | |||||||||||||||||||||
Circulation capacity pressure | 0.35 bar | |||||||||||||||||||||
Usable bath opening (W x L / D) |
18 x 13 / 15 cm | |||||||||||||||||||||
Filling volume | 5 ... 7.5 l | |||||||||||||||||||||
Dimensions
(W × L × H) |
33 x 47 x 70 cm |
Product Features
- Models for internal and external applications
- Bright, white, easy to read display
- Very quiet
- USB connection
- Integrated pump connection M16×1
- easy switching between internal and external circulation
- Bath lid and drain tap included
- removable ventilation grid
- Refrigeration unit without side vents
- Class III (FL) according to DIN 12876-1
- Space-saving cooling coil design yields more usable space in the bath tank
Technical data for CORIO CD-800F
Dimensions (W × L × H) |
33 x 47 x 70 cm |
---|---|
Filling volume | 5 ... 7.5 l |
Weight | 42 kg |
Pump connections | M16x1 male |
Barbed fittings inner diameter | 8/12 mm |
Working temperature range | -40 ... +150 °C |
---|---|
Temperature stability | ±0.03 °C |
Setting the resolution of the temperature display | 0.1 °C |
Temperature display resolution | 0.01 ... 0.1 °C |
Ambient temperature | +5 ... +40 °C |
Heating capacity | 2 kW | |||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
||||||||||||||||||||||
Pump capacity flow pressure | 15 l/min | |||||||||||||||||||||
Circulation capacity pressure | 0.35 bar | |||||||||||||||||||||
Viscosity max. | 50 cSt | |||||||||||||||||||||
Power consumption | 11 A |
Bath tank | Stainless steel |
---|---|
Bath cover | integrated |
Usable bath opening (W x L / D) |
18 x 13 / 15 cm |
Cooling of compressor | 1-stage Air |
---|
Classification according to DIN12876-1 |
Classification III (FL) |
---|---|
IP Code according to IEC 60529 |
IP 21 |
Pump function | Pressure Pump |
Pump type | Immersion Pump |
Interfaces | USB |
---|---|
Temperature control | PID1 |
Absolute temperature calibration | 1 Point Calibration |
Temperature display | LED |
Temperature setting | Keypad |
9012715.N1.13 | 230V/60Hz (Nema N6-20 Plug) (R1270) |
---|---|
9012715.N1.03.chn | 230V/50Hz (CN Plug) (R1270) |
9012715.N1.04 | 230V/50Hz (UK Plug Type BS1363A) (R1270) |
9012715.N1.05 | 230V/50Hz (CH Plug Type SEV 1011) (R1270) |
9012715.N1.02 | 115V/60Hz (Nema N5-20 Plug) (R1270) |
9012715.N1.03 | 230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R1270) |
9012715.N1.12 | 200V/50-60Hz (Nema N6-20 Plug) (R1270) |
9012715.N1.01 | 100V/50-60Hz (Nema N5-20 Plug) (R1270) |
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
Accessory for CORIO CD-800F
Downloads for CORIO CD-800F
We would also be happy to send you our printed catalogs and brochures by post.
Request brochureCORIO CD-800F Benefits
Brilliant.
Very bright display makes it easy to read even from a distance.
Switch on. And off you go.
Intelligent operating concept. Ready for operation with just a few quick and easy steps.
Timer. Integrated.
CORIO circulators include an integrated timer function. When the set time has elapsed, a signal sounds and the device switches off. Setting range: 0 ... 999 minutes.
Everything at the front.
All operating controls and safety functions are accessed easily and comfortably from the front.
Related Products
Products with similar properties