CORIO CD-450F Refrigerated / heating circulator

The new generation of laboratory circulators for routine work and standard tasks

Refrigerated / heating circulators of the CORIO CD series distinguish themselves with a great price-to-performance ratio. Durable and dependable, they are suitable for a wide range of standard and routine work in both laboratories and industrial environments.

Alternatively with natural refrigerant

The CORIO CD-450F is available with natural refrigerant. Order No. 9 012 714.N1

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CD-450F
Order No. see refrigerant variants
Working temperature range -30 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Cooling capacity
Refrigerant variant N1 (Environmentally-friendly natural refrigerant)
Order No. 9012714.N1.04
Medium: Ethanol
°C 20 0 -10 -20 -30
kW 1 0.45 0.38 0.28 0.17 0.07
Refrigerant R290
Global Warming Potential 3
Filling weight 33 g
Carbon dioxide equivalent 9.9E-5 t
Refrigerant variant S1
Order No. 9012714.S1.04
Medium: Ethanol
°C 20 0 -10 -20 -30
kW 1 0.45 0.38 0.28 0.17 0.07
Refrigerant R449A
Global Warming Potential 1397
Filling weight 70 g
Carbon dioxide equivalent 0.09779 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow pressure 15 l/min
Circulation capacity pressure 0.35 bar
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Filling volume 3 ... 4 l
Dimensions
(W × L × H)
23 x 40 x 65 cm
Arrange callback

Product Features

  • Models for internal and external applications
  • Bright, white, easy to read display
  • Very quiet
  • USB connection
  • Integrated pump connection M16×1
  • easy switching between internal and external circulation
  • Bath lid and drain tap included
  • removable ventilation grid
  • Refrigeration unit without side vents
  • Class III (FL) according to DIN 12876-1
  • Space-saving cooling coil design yields more usable space in the bath tank

Technical data for CORIO CD-450F

Dimensions and volumes
Dimensions
(W × L × H)
23 x 40 x 65 cm
Filling volume 3 ... 4 l
Weight 25.1 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -30 ... +150 °C
Temperature stability ±0.03 °C
Setting the resolution of the temperature display 0.1 °C
Temperature display resolution 0.01 ... 0.1 °C
Ambient temperature +5 ... +35 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 2 kW
Cooling capacity
Refrigerant variant N1 (Environmentally-friendly natural refrigerant)
Order No. 9012714.N1.04
Medium: Ethanol
°C 20 0 -10 -20 -30
kW 1 0.45 0.38 0.28 0.17 0.07
Refrigerant R290
Global Warming Potential 3
Filling weight 33 g
Carbon dioxide equivalent 9.9E-5 t
Refrigerant variant S1
Order No. 9012714.S1.04
Medium: Ethanol
°C 20 0 -10 -20 -30
kW 1 0.45 0.38 0.28 0.17 0.07
Refrigerant R449A
Global Warming Potential 1397
Filling weight 70 g
Carbon dioxide equivalent 0.09779 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow pressure 15 l/min
Circulation capacity pressure 0.35 bar
Viscosity max. 50 cSt
Power consumption 13 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Cooling
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Interfaces USB
Temperature control PID1
Absolute temperature calibration 1 Point Calibration
Temperature display LED
Temperature setting Keypad
Available voltage versions
9012714.N1.01 100V/50-60Hz (Nema N5-15 Plug) (R290)
9012714.S1.01 100V/50-60Hz (Nema N5-15 Plug) (R449A)
9012714.N1.02 115V/60Hz (Nema N5-15 Plug) (R290)
9012714.S1.02 115V/60Hz (Nema N5-15 Plug) (R449A)
9012714.N1.03 230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R290)
9012714.S1.03 230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A)
9012714.N1.03.chn 230V/50Hz (CN Plug) (R290)
9012714.S1.03.chn 230V/50Hz (CN Plug) (R449A)
9012714.N1.04 230V/50Hz (UK Plug Type BS1363A) (R290)
9012714.S1.04 230V/50Hz (UK Plug Type BS1363A) (R449A)
9012714.N1.05 230V/50Hz (CH Plug Type SEV 1011) (R290)
9012714.S1.05 230V/50Hz (CH Plug Type SEV 1011) (R449A)
9012714.N1.13 230V/60Hz (Nema N6-20 Plug) (R290)
9012714.S1.13 230V/60Hz (Nema N6-20 Plug) (R449A)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Accessory for CORIO CD-450F

Downloads for CORIO CD-450F

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

CORIO CD-450F Benefits

Brilliant.

Very bright display makes it easy to read even from a distance.

Switch on. And off you go.

Intelligent operating concept. Ready for operation with just a few quick and easy steps.

Timer. Integrated.

CORIO circulators include an integrated timer function. When the set time has elapsed, a signal sounds and the device switches off. Setting range: 0 ... 999 minutes.

Everything at the front.

All operating controls and safety functions are accessed easily and comfortably from the front.

Show all 15 benefits