Print

SE-26 Heating Circulator

Heating Circulator with stainless-steel bath tank for internal and external temperature applications

JULABO Heating Circulators with stainless steel bath tanks are mainly used for internal and external temperature tasks. The new range of these models now have increased capacities, allowing for temperature application of larger external systems, as well as open systems. Simultaneous temperature applications of smaller objects can also be carried out directly in the internal bath of the circulator. Models from the TopTech and HighTech Series’ are available with different bath sizes and features. Accessories for these units include test tube racks, bath covers etc. An integrated cooling coil is provided for temperature applications at near ambient or just below.

Your advantages

  • VFD COMFORT DISPLAY
  • Keypad for setpoints, warning/safety values and menu functions
  • ICC (Intelligent Cascade Control), self-optimizing temperature control
  • ATC3 3-Point-Calibration
  • Pt100 External sensor connection for measurement and control
  • SMART PUMP, electronically adjustable pump stages
  • Early warning system for low liquid level (DBGM 203 06 059.8)
  • Adjustable high temperature cut-out, visible via display
  • RS232 interface for online communication
  • Integrated programmer for 10 program steps

Technical Data

Order No. 9252526
Model series HighTech
Category Heating Circulators
Working temperature range (°C) +20 ... +300
Temperature stability (°C) ±0.01
Setting / display resolution 0.01 °C
Integrated programmer 1x10 steps
Temperature Display VFD
Heating capacity (kW) 3
Pump capacity flow rate (l/min) 22-26
Pump capacity flow pressure (psi) 5.8-10.2
Pump capacity flow suction (psi) 2.9-5.8
Bath opening / bath depth (W x L / D inch) 8.7 x 11.8 / 7.9
Pump connections M16x1
Barbed fittings diameter (inner dia. / mm) 8 / 12
Filling volume (liters) 26
External Pt100 sensor connection integrated
Digital interfaces RS232
Optional Profibus
Ambient temperature 5...40 °C
Dimensions W x L x H (inch) 14.2 x 24 x 18.5
Weight (LBS) 59.4
Classification according to DIN12876-1 Classification III (FL)
Description Built-in cooling coil for tap water connection for temperature applications below or near ambient temperatures are integrated.
Included with each unit 2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 male).
Cooling coil integrated
Bath tank Stainless steel
Power requirement V / Hz / A 230/50-60/13 Additional voltage versions available

Characteristics

Display
A perfect view
A perfect view
Ample, easy to read VFD Comfort display for simultaneous display of 3 values, warning functions, high temperature cut-off, pump stages (resolution 0.01 °C)
Pump stage and liquid level
Pump stage and liquid level
Backlit indicator for selected pump stages and filling volume on Presto® PLUS, Magnum 91 & Forte HT
Operation 
Comfortable and detailed
Comfortable and detailed
Comfortable keypad with additional menu functions for pump stages, calibration, control parameters, programmer, warnings, etc.
Temperature Control
For perfect results
For perfect results
Intelligent Cascade Control’, automatic & self optimizing adjustment of PID control parameters, temperature stability ±0.005 °C internal, <±0.05 °C external
Full control
Full control
Temperature Control Features’ for individual optimization, access to all important control parameters, additional settings for band limit, limits, Co-Speedfactor etc.
Highest measuring accuracy
Highest measuring accuracy
Absolute Temperature Calibration’ for manual compensation of a temperature difference, 3-point calibration
Technical Features
Clever pump system
Clever pump system
Reliable and consistent pump capacity, electronically adjustable pump stages
Control from the external application
Control from the external application
External Pt100 sensor connection for precise measurement and control directly in the external application
Serial connection
Serial connection
RS232 interface for PC connection, e.g. for data communication and recording of measured values
Easy program control
Easy program control
Integrated programmer for the execution of time and temperature dependant profiles, 1 temperature profile with 10 steps max., with real time clock
Warning & Safety Functions
Early warning system for low liquid level
Early warning system for low liquid level
Maximum safety for applications, optical and audible alarm, allows user to refill bath fluid before the unit shuts down
Early warning system for high/low temperature limits
Early warning system for high/low temperature limits
Maximum safety for applications, optical and audible alarm, convertible to automated cut-off function
Enhanced protective functions 
Enhanced protective functions 
Maximum safety, adjustable high temperature cut-off or dry-running protection, additional display of setpoints permits easy and precise adjustments
For flammable bath fluids
For flammable bath fluids
Classification III (FL) according to DIN 12876-1

Other characteristics

For temperature applications near or below ambient temperature: use a cooling coil or JULABO immersion cooler.

Technical changes without prior notification.
Images may deviate from the original.

We would be pleased to submit a non-binding offer for the selected product or answer all pertinent questions.

click here for quotation request SE-26
 

Contact

JULABO USA Inc.
884 Marcon Boulevard
Allentown, PA 18109
Phone: 610-231-0250
Phone: 800-458-5226
(Toll-Free)
sales@julabo.us
julabo.us

Contact form

Available voltage versions

Model: SE-26
Order No.: 9252526
230 V / 50-60 Hz
Catalog request
Order our brochures!