Печать

Booster Pump 8810020

The JULABO magnetically coupled Booster Pump is the ideal solution to increase the pressure or flow rate in your application. The Booster Pump is specifically designed to be easily connected between the PRESTO® and your application The Mag Drive Booster Pump can increase your fluid pressure up to 2.1 bar. The stainless steel design of the Pump provides excellent resistivity against chemical effects. The magnetically driven design guarantees 100% leakage free operation over an extraordinary temperature range of -90 … +250°C.

JULABO Model Booster Pump (magnetically coupled), 2.1 bar
JULABO Order number 8810020
Working Temperature Range -90 … 250 °C
Pump type Magnetically coupled centrifugal pump
Pump material Stainless steel
Suitable fluids Water-Glycol, Silicon oil, Fluorinert®
Maximum Viscosity 50 cSt
Maximum flow rate 80 lpm
Maximum Pressure 2.1 bar*1
Fluid connectors M30x1.5 male*2
Mains power supply 208 – 230V ±10% / 50-60 Hz 1~
Current consumption 1.85 A (208V) / 2A (230V)
Heat input 230W at full motor speed
Control input 3 pin connector to connect to PRESTO® (alarm out). Pump control cable included (5m)
Pump pressure adjustment Manually
Pump pressure indication 2 pressure gauges, for input pressure and ouput pressure Overall Dimensions
Housing Material Stainless steel
Weight (kg) 13.2
Suitable for PRESTO® A40, W40, A45, A80, W80, W91, W92
Номер для заказа: 8810020

Контакты

JULABO GmbH
Gerhard-Juchheim-Strasse 1
77960 Seelbach / Germany
Tel: +49 7823 51-0
info.de@julabo.com
www.julabo.com
Контактный формуляр

Catalog request
Закажите наши брошюры!