DYNEO DD-1001F-BF Beer forcing test refrigerated / heating circulating bath

to determine the ‘best before’ date of beer

The JULABO beer forcing test bath in conjunction with a photometer determines the product life of beer before clouding. The simulated aging process is achieved through a programmable temperature profile, which is repeated, until the first clouding develops.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
DD-1001F-BF
Order No. 9021709.33.D
Working temperature range -38 ... +100 °C
Temperature stability ±0.01 °C
Heating capacity 2 kW
Cooling capacity
Medium: Ethanol
°C 20 10 0 -10 -20 -30
kW* 1
0.95
0.85
0.6
0.32
0.12
Refrigerant R449A
Global Warming Potential 1397
Filling weight 170 g
Carbon dioxide equivalent 0.237 t
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate 8 ... 27 l/min
Pump capacity flow pressure 0.1 ... 0.7 bar
Usable bath opening
(W x L / D)
35 x 41 / 30 cm
Filling volume 42 ... 56 l
Dimensions
(W × L × H)
45 x 64 x 95 cm
Arrange callback Accessory

Product Features

  • Automatic cycles of temperature ramps simulate aging
  • Pre-programmed temperature profiles for forcing test
  • Large bath opening with insert for 20 bottles, 0.5 liters each (Racks for other bottle sizes on request)
  • Powerful and infinitely adjustable pressure pump
  • Flow rate 27 l/min, pressure 0.7 bar
  • Integrated programmer
  • Integrated external Pt100 connection
  • Central rotary knob (controller) simplifies operation
  • Large color TFT display, multilingual interface
Dimensions and volumes
Dimensions
(W × L × H)
45 x 64 x 95 cm
Filling volume 42 ... 56 l
Weight 73.7 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -38 ... +100 °C
Temperature stability ±0.01 °C
Setting the resolution of the temperature display 0.01 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 2 kW
Cooling capacity
Medium: Ethanol
°C 20 10 0 -10 -20 -30
kW* 1
0.95
0.85
0.6
0.32
0.12
Refrigerant R449A
Global Warming Potential 1397
Filling weight 170 g
Carbon dioxide equivalent 0.237 t
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate 8 ... 27 l/min
Pump capacity flow pressure 0.1 ... 0.7 bar
Viscosity max. 50 cST
Power 16 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
35 x 41 / 30 cm
Cooling
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
Pump function Pressure Pump
Pump type Immersion Pump
User Interface Language Chinese / English / French / German / Italian / Japanese / Korean / Portuguese / Russian / Spanish
Electronics
Interfaces Alarm output optional / RS232 optional / Reg/Eprog optional / Standby-Input optional / USB
External pt100 sensor connection integrated
Integrated programmer 8x60 steps
Temperature control PID3
Absolute temperature calibration 3 Point Calibration
Temperature display 3.5" TFT Display
Temperature setting Shaft Encoder
Electronic Timer
hr:min
00:00 ... 99:59
Available voltage versions
9021709.04.D 230V/50-60Hz (UK Plug Type BS1363A) (R449A)
9021709.05.D 230V/50-60Hz (CH Plug Type SEV 1011) (R449A)
9021709.33.D 230V/50-60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A)
9021709.33.chn.D 230V/50-60Hz (CN Plug) (R449A)
Included in delivery
Stainless steel insert basket for 20 x 0.5 liter bottles Plexiglass bath cover.

Accessory for DYNEO DD-1001F-BF

Contact expert

We would also be happy to send you our printed catalogs and brochures by post.


Request brochure

Benefits

  • more-bath
    More bath.
    Designed for more comfort. Thanks to the recessed cooling coil, the internal bath provides more space.
  • space-saving
    Space saving. Free up space.
    Place your JULABO Circulator right next to an application, another unit, or wall. That saves space. This is made possible by eliminating vents and connections on the sides.
  • Solid
    Solid.
    Minimized energy loss through high-quality insulation.
  • drain-tap
    Tidy.
    The special drain tap for easy draining of bath fluids without tools.