FP51-SL Ultra-Low Refrigerated / Heating Circulator

low temperatures below -50 °C with bath opening for internal and external applications

The Ultra-Low Refrigerated Circulators of the HighTech Series with HL or SL circulator feature powerful pressure and suction pumps. The in-struments provide the entire variety of functions of the professional HighTech Series of circulators.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
FP51-SL
Order No. see refrigerant variants
Working temperature range -51 ... +200 °C
Temperature stability ±0.05 °C
Heating capacity 3 kW
Cooling capacity
Refrigerant variant
Order No. 9352751.07
Medium: Ethanol
°C 20 0 -20 -40
kW 1 2 1.5 1 0.26
Refrigerant R507
Global Warming Potential 3985
Filling weight 1030 g
Carbon dioxide equivalent 4.10455 t
Refrigerant variant S1
Order No. 9352751.S1.07
Medium: Ethanol
°C 20 0 -20 -40
kW 1 2 1.5 1 0.26
Refrigerant R449A
Global Warming Potential 1397
Filling weight 1030 g
Carbon dioxide equivalent 1.43891 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Usable bath opening
(W x L / D)
18 x 12 / 20 cm
Filling volume 7.5 ... 12 l
Dimensions
(W × L × H)
46 x 55 x 89 cm
Arrange callback

Product Features

  • VFD COMFORT DISPLAY
  • LCD DIALOG DISPLAY backlit for convenient interactive operation
  • Keypad for setpoints, warning/safety values and menu functions
  • ICC (Intelligent Cascade Control), self-optimizing temperature control
  • TCF Temperature Control Features to optimize the control behavior
  • ATC3 3-Point-Calibration
  • Pt100 External sensor connection for measurement and control
  • SMART PUMP, electronically adjustable pump stages
  • Unique early warning system for low liquid level
  • Adjustable high temperature cut-out, visible via display
  • RS232/RS485 interface for online communication
  • Integrated programmer for 6 x 60 program steps
  • Connections for solenoid valve and HSP booster pump
  • Proportional cooling control
  • Active Cooling Control

Technical data for FP51-SL

Dimensions and volumes
Dimensions
(W × L × H)
46 x 55 x 89 cm
Filling volume 7.5 ... 12 l
Weight 90 kg
Pump connections M16x1 male
Temperature values
Working temperature range -51 ... +200 °C
Temperature stability ±0.05 °C
Setting the resolution of the temperature display 0.01 °C
Setting the resolution of the temperature display 0.01 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 3 kW
Cooling capacity
Refrigerant variant
Order No. 9352751.07
Medium: Ethanol
°C 20 0 -20 -40
kW 1 2 1.5 1 0.26
Refrigerant R507
Global Warming Potential 3985
Filling weight 1030 g
Carbon dioxide equivalent 4.10455 t
Refrigerant variant S1
Order No. 9352751.S1.07
Medium: Ethanol
°C 20 0 -20 -40
kW 1 2 1.5 1 0.26
Refrigerant R449A
Global Warming Potential 1397
Filling weight 1030 g
Carbon dioxide equivalent 1.43891 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cSt
Power consumption 16 A
Bath
Bath cover integrated
Usable bath opening
(W x L / D)
18 x 12 / 20 cm
Cooling
Cooling of compressor 1-stage Air
Other
Sound pressure level 63 dbA
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Electronics
Interfaces Profibus optional
External pt100 sensor connection integrated
Integrated programmer 6x60 steps
Temperature control ICC
Absolute temperature calibration 3 Point Calibration
Temperature display VFD
Temperature setting Keypad
Available voltage versions
9352751.06 230V/3PPE/50Hz (Without Plug) (R507)
9352751.07 400V/3PNPE/50Hz (Plug 16A CEE) (R507)
9352751.S1.07 400V/3PNPE/50Hz (Plug 16A CEE) (R449A)
9352751.16 208V/3PPE/60Hz (Without Plug) (R507)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)
Contact expert

Cool-down time

Medium: Ethanol

Heat-up time

Medium: Thermal

Pump capacity

Medium: Water

Accessory for FP51-SL

Downloads for FP51-SL

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

FP51-SL Benefits

For flammable bath fluid

Classification III (FL) according to DIN 12876-1

ATC3. Calibration.

‘Absolute Temperature Calibration’ for compensating a physically caused temperature difference, 3-point calibration.

Process. Under control.

Full regulation of the dynamics control, access to all important control parameters for individual process optimization.

100 % Cooling capacity

‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures

Show all 17 benefits