FP89-ME TopTech ME Ultra-Low Refrigerated / Heating Circulator

Ultra-low Refrigerated Circulator with 2-stage cascade-cooling

The Ultra-Low Refrigerated Circulators of the TopTech Series are equipped with a dual-stage cascade refrigeration system for continuous operation of internal and external temperature applications.

Models with ME circulator

  • Heated bath cover plate to prevent condensation or ice build-up
  • Pressure pump up to 0.45 bar, electronically adjustable in steps
  • ACC Active Cooling Control across the entire temperature range
  • Compact design Note: FP models feature an energy-saving proportional cooling control.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
FP89-ME
Order No. 9162689.13
Working temperature range -90 ... +100 °C
Temperature stability ±0.02 °C
Heating capacity 1.2 kW
Cooling capacity
Medium: Ethanol
°C 20 0 -20 -40 -60 -80
kW 1 1 0.92 0.88 0.75 0.58 0.1
Refrigerant Information Refrigerant stage 1 Refrigerant stage 2
Refrigerant R404A R23
Global Warming Potential 3922 14800
Filling weight 500 g 200 g
Carbon dioxide equivalent 1.961 t 2.96 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow rate 11 ... 16 l/min
Pump capacity flow pressure 3.3 ... 6.5 psi
Usable bath opening
(W x L / D)
5.1 x 5.9 / 6.3 in.
Filling volume 5.5 ... 8 l
Dimensions
(W × L × H)
21.7 x 23.6 x 35.4 in.
Arrange callback

Product Features

  • PID3 cascade temperature control
  • RS232 interface for online communication
  • Integrated programmer for 10 program steps
  • VFD COMFORT DISPLAY
  • Keypad for setpoints, warning/safety values and menu functions
  • ATC3 3-Point-Calibration
  • Pt100 External sensor connection for measurement and control
  • SMART PUMP, electronically adjustable pump stages
  • Adjustable high temperature cut-out, visible via display
  • Active Cooling Control
  • Unique early warning system for low liquid level

Technical data for FP89-ME

Dimensions and volumes
Dimensions
(W × L × H)
21.7 x 23.6 x 35.4 in.
Filling volume 5.5 ... 8 l
Weight 302 lbs
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -90 ... +100 °C
Temperature stability ±0.02 °C
Setting the resolution of the temperature display 0.01 °C
Setting the resolution of the temperature display 0.01 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 1.2 kW
Cooling capacity
Medium: Ethanol
°C 20 0 -20 -40 -60 -80
kW 1 1 0.92 0.88 0.75 0.58 0.1
Refrigerant Information Refrigerant stage 1 Refrigerant stage 2
Refrigerant R404A R23
Global Warming Potential 3922 14800
Filling weight 500 g 200 g
Carbon dioxide equivalent 1.961 t 2.96 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow rate 11 ... 16 l/min
Pump capacity flow pressure 3.3 ... 6.5 psi
Viscosity max. 50 cSt
Power consumption 16 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
5.1 x 5.9 / 6.3 in.
Cooling
Cooling of compressor 2-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump type Immersion Pump
Electronics
Interfaces Alarm output / RS232
External pt100 sensor connection integrated
Integrated programmer 1x10 steps
Temperature control PID3
Absolute temperature calibration 3 Point Calibration
Temperature display VFD
Temperature setting Keypad
Available voltage versions
9162689.03 230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R404A)
9162689.13 230V/60Hz (Nema N6-20 Plug) (R404A)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Accessory for FP89-ME

Downloads for FP89-ME

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

FP89-ME Benefits

For flammable bath fluid

Classification III (FL) according to DIN 12876-1

ATC3. Calibration.

‘Absolute Temperature Calibration’ for compensating a physically caused temperature difference, 3-point calibration.

100 % Cooling capacity

‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures

Early warning system for high/low temperature limits

Maximum safety for applications, optical and audible alarm, convertible to automated cut-off function

Show all 14 benefits