CORIO CP-1000F Refrigerated / heating circulator

with an enhanced pump performance and a temperature range up to +200 °C

Refrigerated circulators from the CORIO CP range are suitable for applications with a temperature range up to +200°C. The enhanced pump performance ensures they are suitable for easy temperature control tasks in combination with external applications.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CP-1000F
Order No. 9013707.02
Working temperature range -50 ... +200 °C
Temperature stability ±0.03 °C
Heating capacity 1 kW
Cooling capacity
Medium: Ethanol
°C 20 10 0 -10 -20 -30 -40
kW 1 1 1 0.96 0.73 0.51 0.25 0.11
Refrigerant R449A
Global Warming Potential 1397
Filling weight 190 g
Carbon dioxide equivalent 0.26543 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow pressure 8 ... 27 l/min
Circulation capacity pressure 1.5 ... 10.2 psi
Usable bath opening
(W x L / D)
7.1 x 5.1 / 5.9 in.
Filling volume 5 ... 7.5 l
Dimensions
(W × L × H)
16.5 x 19.3 x 29.1 in.
Arrange callback

Product Features

  • Models for internal and external applications
  • Bright, white, easy to read display
  • Very quiet
  • powerful and infinitely adjustable pressure pump
  • USB connection
  • RS232 interface for online communication
  • Bath lid and drain tap included
  • Removable ventilation grid
  • Refrigeration unit without side vents
  • Class III (FL) according to DIN 12876-1
  • easy switching between internal and external circulation
  • Integrated pump connection M16×1
  • Space-saving cooling coil design provides more usable space in the bath tank
  • early warning system for low liquid level

Technical data for CORIO CP-1000F

Dimensions and volumes
Dimensions
(W × L × H)
16.5 x 19.3 x 29.1 in.
Filling volume 5 ... 7.5 l
Weight 113.5 lbs
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -50 ... +200 °C
Temperature stability ±0.03 °C
Setting the resolution of the temperature display 0.1 °C
Temperature display resolution 0.01 ... 0.1 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 1 kW
Cooling capacity
Medium: Ethanol
°C 20 10 0 -10 -20 -30 -40
kW 1 1 1 0.96 0.73 0.51 0.25 0.11
Refrigerant R449A
Global Warming Potential 1397
Filling weight 190 g
Carbon dioxide equivalent 0.26543 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow pressure 8 ... 27 l/min
Circulation capacity pressure 1.5 ... 10.2 psi
Viscosity max. 50 cSt
Power consumption 16 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
7.1 x 5.1 / 5.9 in.
Cooling
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Interfaces RS232 / USB
Temperature control PID1
Absolute temperature calibration 3 Point Calibration
Temperature display LED
Temperature setting Keypad
Electronic Timer
h:min
0 ... 999
Available voltage versions
9013707.02 115V/60Hz (Nema N5-20 Plug) (R449A)
9013707.04 200-230V/50-60Hz (UK Plug Type BS1363A) (R449A)
9013707.05 200-230V/50-60Hz (CH Plug Type SEV 1011) (R449A)
9013707.33 200-230V/50-60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A)
9013707.33.chn 200-230V/50-60Hz (CN Plug) (R449A)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)
Contact expert

Cool-down time

Medium: Ethanol

Pump capacity

Medium: Water

Accessory for CORIO CP-1000F

Downloads for CORIO CP-1000F

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

CORIO CP-1000F Benefits

Brilliant.

Very bright display makes it easy to read even from a distance.

Switch on. And off you go.

Intelligent operating concept. Ready for operation with just a few quick and easy steps.

Timer. Integrated.

CORIO circulators include an integrated timer function. When the set time has elapsed, a signal sounds and the device switches off. Setting range: 0 ... 999 minutes.

Everything at the front.

All operating controls and safety functions are accessed easily and comfortably from the front.

Show all 28 benefits