PRESTO W85t Process system
Powerhouses among the process system
High cooling capacities enable extremely low temperatures down to -85 °C possible. The high heating capacity, particularly with the A85t and the W85t, provides even more flexibility in the application.
The highly dynamic temperature control systems PRESTO are designed for high-precision temperature control for a wide range of applications such as reactor vessels or material stress tests. Moreover, by using efficient components, the process systems can compensate exothermic and endothermic reactions exceptionally fast. Permanent internal monitoring and self-lubricating pumps ensure a long life-time. In addition, numerous interfaces offer many remote control possibilities across networks or for integration into higher-level control systems.
Order No. | 9421852.16.T | ||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -85 ... +250 °C | ||||||||||||||||||||||||||||||||||
Temperature stability | ±0.05 ... ±0.1 °C | ||||||||||||||||||||||||||||||||||
Heating capacity | 12.5 kW | ||||||||||||||||||||||||||||||||||
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
|||||||||||||||||||||||||||||||||||
Pump capacity flow pressure | 0 ... 80 l/min | ||||||||||||||||||||||||||||||||||
Circulation capacity pressure | 1.5 ... 43.5 psi | ||||||||||||||||||||||||||||||||||
Dimensions
(W × L × H) |
24 x 33.3 x 49.2 in. |
Product Features
- Cooling capacity up to 2.8 kW
- External Pt100 sensor connection
- Connections for alarm-output, Pt100 external sensor and standby-input
- Analog connections, RS485, Profibus DP (accessory)
- Built-in 5.7’’ industrial color touchscreen
- Second external Pt100 sensor connection (accessory)
- Alarm output
- Heating capacity up to 15 kW
- Temperature stability ±0.05 °C ... ±0.1 °C
- Pump pressure up to 3 bar, max. flow rate 80 l/min
Technical data for PRESTO W85t
Dimensions (W × L × H) |
24 x 33.3 x 49.2 in. |
---|---|
Internal usable expansion volume | 7 l |
Minimal process volume | 9.5 l |
Active heat exchanger volume | 5 l |
Weight | 738.5 lbs |
Pump connections | M30x1.5 male |
Cooling Water Connection | G¾ in. |
Working temperature range | -85 ... +250 °C |
---|---|
Temperature stability | ±0.05 ... ±0.1 °C |
Setting the resolution of the temperature display | 0.01 °C |
Temperature display resolution | 0.01 °C |
Ambient temperature | +5 ... +40 °C |
Heating capacity | 12.5 kW | ||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
|||||||||||||||||||||||||||||||||||
Pump capacity flow pressure | 0 ... 80 l/min | ||||||||||||||||||||||||||||||||||
Circulation capacity pressure | 1.5 ... 43.5 psi | ||||||||||||||||||||||||||||||||||
Viscosity max. | 50 cSt | ||||||||||||||||||||||||||||||||||
Power consumption | 55 A |
Cooling of compressor | 2-stage Water |
---|---|
Cooling water pressure max. | 87 psi |
Max. heat dissipated by unit into cooling water | 7 kW |
to provide full cooling capacity according to catalog values | |
Cooling water temperature range | 10 ... 15 °C |
---|---|
Cooling water difference pressure | 21.8 ... 87 psi |
Cooling water consumption 2 | 6 l/min |
2 Cooling water consumption may vary outside recommended cooling water properties. |
Cooling water temperature range | 5 ... 35 °C |
---|---|
Cooling water difference pressure 3 | 7.3 ... 87 psi |
3 At cooling water temperatures of 25 °C and higher, the minimum differential pressure is 1 bar. |
Sound pressure level | 69 dbA |
---|---|
Classification according to DIN12876-1 |
Classification III (FL) |
IP Code according to IEC 60529 |
IP 20 |
Pump type | Centrifugal Pump (Magnetically coupled) |
Interfaces | Alarm output / Ethernet / Modbus / Profibus optional / REG/EPROG optional / RS232 / RS485 optional / SD memory card / Standby-Input optional / USB |
---|---|
External pt100 sensor connection | integrated |
2nd external Pt100 sensor connection | accessory |
Integrated programmer | 8x60 steps |
Temperature control | ICC |
Absolute temperature calibration | 3 Point Calibration |
Temperature display | 5.7" TFT Touchscreen |
Temperature setting | Touchscreen |
9421852.07.T | 400V/3PNPE/50Hz (Plug 32A CEE) (R507) |
---|---|
9421852.16.T | 208-230V/3PPE/60Hz (Without Plug) (R507) |
9421852.06.T | 230V/3PPE/50Hz (Plug 63A CEE) (R507) |
Accessory for PRESTO W85t
Downloads for PRESTO W85t
We would also be happy to send you our printed catalogs and brochures by post.
Request brochurePRESTO W85t Benefits
Touch display. Perfect operation.
With the touch display, the user always has an overview of all values and functions. The intuitive and multilingual menu structure enables perfect control.
Convenience for several users
Administrator level for customizing instrument settings, user levels with limited permissions for fast and safe defined access, password protection, all levels adjustable
100 % Cooling capacity
‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures
Intelligent temperature control.
Intelligent cascade control - automatic and self-optimizing adaptation of the PID control parameters with external stability of +/- 0.05 °C.