F70-ME Ultra-low refrigerated / heating circulator

Ultra-low refrigerated circulator with 2-stage cascade-cooling

The ultra-low refrigerated circulators of the TopTech series are equipped with a dual-stage cascade refrigeration system for continuous operation of internal and external temperature applications.

Models with ME circulator

  • Heated bath cover plate to prevent condensation or ice build-up
  • Pressure pump up to 0.45 bar, electronically adjustable in steps
  • ACC Active Cooling Control across the entire temperature range
  • Compact design Note: FP models feature an energy-saving proportional cooling control.
Data at mains voltage / frequency:
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F)
F70-ME
Order No. 9 162 670
Working temperature range -70 ... +100 °C
Temperature stability +/-0.02 °C
Heating capacity 1.2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -40 -60
0.34 0.22 0.17 0.13 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -40 -60
kW 0.34 0.22 0.17 0.13 0.07
Pump capacity flow rate 11 ... 16 l/min
Pump capacity flow pressure 0.23 ... 0.45 bar
Usable bath opening
(W x L / D)
12 x 12 / 13 cm
Filling volume 3.5 ... 4.5 l
Dimensions
(W × L × H)
42 x 54 x 71 cm
Arrange callback Accessory

Product Features

  • PID3 cascade temperature control
  • RS232 interface for online communication
  • Integrated programmer for 10 program steps
Dimensions and volumes
Dimensions
(W × L × H)
42 x 54 x 71 cm
Filling volume 3.5 ... 4.5 l
Weight 63 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -70 ... +100 °C
Temperature stability +/-0.02 °C
Setting the resolution of the temperature display 0.01 °C
Temperature display resolution 0.01 °C
Ambient temperature +5.0 ... +40.0 °C
Data at mains voltage / frequency:
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F)
Performance values
Heating capacity 1.2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -40 -60
0.34 0.22 0.17 0.13 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -40 -60
kW 0.34 0.22 0.17 0.13 0.07
Pump capacity flow rate 11 ... 16 l/min
Pump capacity flow pressure 0.23 ... 0.45 bar
Viscosity max. 50 cST
Refrigerant stage 1
Refrigerant R404A
Global Warming Potential for R404A 3922
Filling volume 300 g
Carbon dioxide equivalent 1.177 t
Power 14 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
12 x 12 / 13 cm
Cooling
Cooling of compressor 2-stage Air
Other
Sound pressure level 56 dbA
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump type Immersion Pump
Electronics
External pt100 sensor connection integrated
Integrated programmer 1x10 steps
Temperature control PID3
Absolute temperature calibration 3 Point Calibration
Temperature display VFD
Temperature setting Keypad
Available voltage versions
230V/50Hz Schuko Plug - CEE 7/4 Plug Type F 9162670.03
Included in delivery
2 barbed fittings suitable for tubing 8 and 10 mm inner dia. (pump connections M10x1 female)

Accessory for F70-ME

Contact expert

We would also be happy to send you our printed catalogs and brochures by post.


Request brochure

Benefits

  • 100% geprüft
    100% Checked.
    100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.
  • Blatt wächst in einer Glühbirne
    Green technology.
    Development consistently applied environmentally friendly materials and technologies.
  • Made in Germany
    JULABO. Quality.
    Highest standards of quality for a long product life.
  • Quickstart
    Quick start.
    Individual JULABO consultation and comprehensive manuals at your disposal.