F32-HL Refrigerated / heating circulator

Top-of-the-line for challenging tasks

The top-of-the-line models with HL circulators offer professional technology and maximum functionality. The additional LCD dialog display interactively assists the user in setting circulator parameters. Intelligent Cascade Control (ICC) and Temperature Control Features (TCF) ensure optimal control behavior. HL circulators also have a Stakei connection.

Data at mains voltage / frequency:
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F)
F32-HL
Order No. 9 312 632
Working temperature range -35 ... +200 °C
Temperature stability ±0.01 °C
Heating capacity 0.8 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Usable bath opening
(W x L / D)
18 x 12 / 15 cm
Filling volume 5.5 ... 8 l
Dimensions
(W × L × H)
31 x 42 x 66 cm
F32-HL
Order No. 9 312 632
Working temperature range -35 ... +200 °C
Temperature stability ±0.01 °C
Heating capacity 0.8 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Usable bath opening
(W x L / D)
18 x 12 / 15 cm
Filling volume 5.5 ... 8 l
Dimensions
(W × L × H)
31 x 42 x 66 cm
F32-HL
Order No. 9 312 632
Working temperature range -35 ... +200 °C
Temperature stability ±0.01 °C
Heating capacity 1 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Usable bath opening
(W x L / D)
18 x 12 / 15 cm
Filling volume 5.5 ... 8 l
Dimensions
(W × L × H)
31 x 42 x 66 cm
F32-HL
Order No. 9 312 632
Working temperature range -35 ... +200 °C
Temperature stability ±0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Usable bath opening
(W x L / D)
18 x 12 / 15 cm
Filling volume 5.5 ... 8 l
Dimensions
(W × L × H)
31 x 42 x 66 cm
F32-HL
Order No. 9 312 632
Working temperature range -35 ... +200 °C
Temperature stability ±0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Usable bath opening
(W x L / D)
18 x 12 / 15 cm
Filling volume 5.5 ... 8 l
Dimensions
(W × L × H)
31 x 42 x 66 cm
F32-HL
Order No. 9 312 632
Working temperature range -35 ... +200 °C
Temperature stability ±0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Usable bath opening
(W x L / D)
18 x 12 / 15 cm
Filling volume 5.5 ... 8 l
Dimensions
(W × L × H)
31 x 42 x 66 cm
F32-HL
Order No. 9 312 632
Working temperature range -35 ... +200 °C
Temperature stability ±0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Usable bath opening
(W x L / D)
18 x 12 / 15 cm
Filling volume 5.5 ... 8 l
Dimensions
(W × L × H)
31 x 42 x 66 cm
F32-HL
Order No. 9 312 632
Working temperature range -35 ... +200 °C
Temperature stability ±0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Usable bath opening
(W x L / D)
18 x 12 / 15 cm
Filling volume 5.5 ... 8 l
Dimensions
(W × L × H)
31 x 42 x 66 cm
F32-HL
Order No. 9 312 632
Working temperature range -35 ... +200 °C
Temperature stability ±0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Usable bath opening
(W x L / D)
18 x 12 / 15 cm
Filling volume 5.5 ... 8 l
Dimensions
(W × L × H)
31 x 42 x 66 cm
Arrange callback Accessory

Product Features

  • VFD COMFORT DISPLAY
  • LCD DIALOG DISPLAY backlit for convenient interactive operation
  • Keypad for setpoints, warning/safety values and menu functions
  • ICC (Intelligent Cascade Control), self-optimizing temperature control
  • TCF Temperature Control Features to optimize the control behavior
  • ATC3 3-Point-Calibration
  • Pt100 External sensor connection for measurement and control
  • SMART PUMP, electronically adjustable pump stages
  • Adjustable pressure and suction pump
  • Adjustable high temperature cut-out, visible via display
  • RS232/RS485 interface for online communication
  • Optional: analogue interfaces
  • Integrated programmer for 6 x 60 program steps
  • Connections for solenoid valve and HSP booster pump
  • Active Cooling Control
Dimensions and volumes
Dimensions
(W × L × H)
31 x 42 x 66 cm
Filling volume 5.5 ... 8 l
Weight 38 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -35 ... +200 °C
Temperature stability ±0.01 °C
Setting the resolution of the temperature display 0.01 °C
Temperature display resolution 0.01 °C
Ambient temperature +5.0 ... +40.0 °C
Data at mains voltage / frequency:
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F)
Performance values
Heating capacity 0.8 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R134a
Global Warming Potential for R134a 1430
Filling volume 166 g
Carbon dioxide equivalent 0.237 t
Power 14 A
Performance values
Heating capacity 0.8 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R134a
Global Warming Potential for R134a 1430
Filling volume 166 g
Carbon dioxide equivalent 0.237 t
Power 14 A
Performance values
Heating capacity 1 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R134a
Global Warming Potential for R134a 1430
Filling volume 166 g
Carbon dioxide equivalent 0.237 t
Power 14 A
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R134a
Global Warming Potential for R134a 1430
Filling volume 166 g
Carbon dioxide equivalent 0.237 t
Power 12 A
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R134a
Global Warming Potential for R134a 1430
Filling volume 166 g
Carbon dioxide equivalent 0.237 t
Power 12 A
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R134a
Global Warming Potential for R134a 1430
Filling volume 166 g
Carbon dioxide equivalent 0.237 t
Power 2 A
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R134a
Global Warming Potential for R134a 1430
Filling volume 166 g
Carbon dioxide equivalent 0.237 t
Power 2 A
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R134a
Global Warming Potential for R134a 1430
Filling volume 166 g
Carbon dioxide equivalent 0.237 t
Power 12 A
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30
0.45 0.39 0.15 0.06
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30
kW 0.45 0.39 0.15 0.06
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R134a
Global Warming Potential for R134a 1430
Filling volume 166 g
Carbon dioxide equivalent 0.237 t
Power 12 A
Bath
Bath cover integrated
Usable bath opening
(W x L / D)
18 x 12 / 15 cm
Cooling
Cooling of compressor 1-stage Air
Other
Sound pressure level 53 dbA
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump type Immersion Pump
Electronics
Digital interface Profibus optional
External pt100 sensor connection integrated
Integrated programmer 6x60 steps
Temperature control ICC
Absolute temperature calibration 3 Point Calibration
Temperature display VFD
Temperature setting Keypad
Available voltage versions
100V/50-60Hz Nema N5-15 Plug 9312632.01
115V/60Hz Nema N5-15 Plug 9312632.02
230V/50-60Hz UK Plug Type BS1363A 9312632.04
230V/50-60Hz CH Plug Type SEV 1011 9312632.05
230V/50-60Hz Schuko Plug - CEE 7/4 Plug Type F 9312632.33
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Accessory for F32-HL

Contact expert

We would also be happy to send you our printed catalogs and brochures by post.


Request brochure

Benefits

  • 100% geprüft
    100% Checked.
    100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.
  • Blatt wächst in einer Glühbirne
    Green technology.
    Development consistently applied environmentally friendly materials and technologies.
  • ICC
    Intelligent temperature control.
    Intelligent cascade control - automatic and self-optimizing adaptation of the PID control parameters with external stability of +/- 0.05 °C.
  • Made in Germany
    JULABO. Quality.
    Highest standards of quality for a long product life.