CF30 Cryo-Compact Circulator

The Cryo-Compact Circulators of the CF Series provide powerful cooling and heating capabilities in a space-saving, compact design. The instru-ments feature 2 kW heating capacity and classification III according to DIN 12876-1. The Cryo-Compact Circulators have pump connections for external temperature control applications and a bath opening for tem-perature control of small objects.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CF30
Order No. 9400330.02
Working temperature range -30 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 1 kW
Cooling capacity
Medium: Ethanol
°C 20 0 -20
kW 1 0.32 0.25 0.15
Refrigerant R134a
Global Warming Potential 1430
Filling weight 150 g
Carbon dioxide equivalent 0.2145 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow pressure 15 l/min
Circulation capacity pressure 5.1 psi
Usable bath opening
(W x L / D)
6.3 x 1.2 / 5.5 in.
Filling volume 2 ... 3.5 l
Dimensions
(W × L × H)
9.4 x 18.1 x 15.7 in.
Arrange callback

Product Features

  • Ergonomic design and easy operation
  • Small overall dimensions for lowest space requirement
  • Splash-proof keypad
  • Large, bright LED display
  • Precise PID temperature control
  • RS232 interface for PC connection
  • Adjustable high temperature cut-out, visible via display
  • Pump connections for external temperature applications
  • Bath opening for temperature control of small objects in the bath
  • Front drain
  • Extra bright displays, easy to read from a distance
  • Precise temperature control
  • Unique early warning system for low liquid level
  • ATC3 3-Point-Calibration
  • Class III (FL) according to DIN 12876-1

Technical data for CF30

Dimensions and volumes
Dimensions
(W × L × H)
9.4 x 18.1 x 15.7 in.
Filling volume 2 ... 3.5 l
Weight 77.2 lbs
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -30 ... +150 °C
Temperature stability ±0.03 °C
Temperature display resolution 0.1 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 1 kW
Cooling capacity
Medium: Ethanol
°C 20 0 -20
kW 1 0.32 0.25 0.15
Refrigerant R134a
Global Warming Potential 1430
Filling weight 150 g
Carbon dioxide equivalent 0.2145 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow pressure 15 l/min
Circulation capacity pressure 5.1 psi
Viscosity max. 30 cSt
Power consumption 13 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
6.3 x 1.2 / 5.5 in.
Cooling
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump type Immersion Pump
Electronics
Temperature control PID1
Temperature display LED
Available voltage versions
9400330.13 230V/60Hz (Nema N6-20 Plug) (R134a)
9400330.02 115V/60Hz (Nema N5-20 Plug) (R134a)
9400330.03 230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R134a)
9400330.04 230V/50Hz (UK Plug Type BS1363A) (R134a)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Accessory for CF30

Downloads for CF30

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

CF30 Benefits

100% Checked.

100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.

Green technology.

Development consistently applied environmentally friendly materials and technologies.

JULABO. Quality.

Highest standards of quality for a long product life.

Quick start.

Individual JULABO consultation and comprehensive manuals at your disposal.

Show all 7 benefits