CORIO CP-1201F Refrigerated / heating circulator new

with an enhanced pump performance and a temperature range up to +200 °C

Refrigerated circulators from the CORIO CP range are suitable for applications with a temperature range up to +100°C. The enhanced pump performance ensures they are suitable for easy temperature control tasks in combination with external applications.

The CORIO CP-1201F refrigerated circulator is exceptionally versatile with a bath opening of 35 x 41 cm (width x depth) and a bath depth of 30 cm. The resulting bath volume of 50 liters makes it possible to control the temperature of large objects, or many smaller objects, directly in the bath.

This cooling machine uses natural, environmentally-friendly refrigerant and was developed with a focus on energy efficiency. This means significant savings of up to 70% with respect to the operating costs for numerous applications, which also means faster amortization of the procurement cost. At the same time, the lower energy consumption positively contributes to climate protection.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CP-1201F
Order number including voltage version 9013718.N1.22
Working temperature range -40 ... 100 °C
Temperature stability ±0.03 °C
Heating capacity 1 kW
Pump capacity flow rate 8 ... 27 l/min
Pump capacity pressure 1.5 ... 10.2 psi
Usable bath opening
(W x L / D)
13.8 x 15.7 / 11.8 in.
Filling volume 48 ... 56 l
Total dimensions
(W × L × H)
17.7 x 25.2 x 37.4 in.
Arrange callback

Product Features

  • Models for internal and external applications
  • Bright, white, easy to read display
  • Very quiet
  • powerful and infinitely adjustable pressure pump
  • USB connection
  • RS232 interface for online communication
  • Bath lid and drain tap included
  • Removable ventilation grid
  • Refrigeration unit without side vents
  • Class III (FL) according to DIN 12876-1
  • easy switching between internal and external circulation
  • Integrated pump connection M16×1
  • Space-saving cooling coil design provides more usable space in the bath tank
  • early warning system for low liquid level

Technical data for CORIO CP-1201F

Dimensions and volumes
Total dimensions
(W × L × H)
17.7 x 25.2 x 37.4 in.
Filling volume 48 ... 56 l
Weight 149.9 lbs
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -40 ... 100 °C
Temperature stability ±0.03 °C
Temperature display resolution 0.01 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 1 kW
Cooling capacity
Medium: Ethanol
°C 20 10 0 -10 -20 -30 -40
kW 1 1.25 1.2 1.1 0.85 0.54 0.3 0.05
Refrigerant R1270
Global Warming Potential 2
Filling weight 90 g
Carbon dioxide equivalent 0.00018 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow rate 8 ... 27 l/min
Pump capacity pressure 1.5 ... 10.2 psi
Viscosity max. 50 cSt
Power consumption 11 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
13.8 x 15.7 / 11.8 in.
Cooling
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Interfaces RS232 / USB
Temperature control PID1
Absolute temperature calibration 3 Point Calibration
Temperature display LED
Temperature setting Keypad
Electronic Timer
h:min
0 ... 999
Available voltage versions
9013718.N1.04 200-230V 50/60Hz (UK Plug Type BS1363A) (R1270)
9013718.N1.05 200-230V 50/60Hz (CH Plug Type T12 / 10A) (R1270)
9013718.N1.33 200-230V 50/60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R1270)
9013718.N1.33.chn 200-230V 50/60Hz (CN Plug) (R1270)
9013718.N1.22 100-115V 50/60Hz (Nema N5-20 Plug) (R1270)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Accessory for CORIO CP-1201F

Downloads for CORIO CP-1201F

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

CORIO CP-1201F Benefits

Brilliant.

Very bright display makes it easy to read even from a distance.

Connection. Easy.

Inclined pump connections (M16×1) facilitate the connection of applications. Each unit includes 2 barbed fittings of 8/12 mm diameter each.

Everything at the front.

All operating controls and safety functions are accessed easily and comfortably from the front.

Exact.

You can rely on it. PID1 control and 'Active Cooling Control' make the new CORIO precise and perfect.

Show all 10 benefits