CORIO CD-1001F Refrigerated / heating circulator
The new generation of laboratory circulators for routine work and standard tasks
Refrigerated / heating circulators of the CORIO CD series distinguish themselves with a great price-to-performance ratio. Durable and dependable, they are suitable for a wide range of standard and routine work in both laboratories and industrial environments.
Order No. | 9012708.13 | |||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -38 ... +100 °C | |||||||||||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||||||||||
Heating capacity | 2 kW | |||||||||||||||||||||||
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
||||||||||||||||||||||||
Pump capacity flow pressure | 17 l/min | |||||||||||||||||||||||
Circulation capacity pressure | 6.2 psi | |||||||||||||||||||||||
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. | |||||||||||||||||||||||
Filling volume | 48 ... 56 l | |||||||||||||||||||||||
Dimensions
(W × L × H) |
17.7 x 25.2 x 37.4 in. |
Product Features
- Models for internal and external applications
- Bright, white, easy to read display
- Very quiet
- USB connection
- Integrated pump connection M16×1
- easy switching between internal and external circulation
- Bath lid and drain tap included
- Removable ventilation grid
- Refrigeration unit without side vents
- Class III (FL) according to DIN 12876-1
- Space-saving cooling coil design provides more usable space in the bath tank
Technical data for CORIO CD-1001F
Dimensions (W × L × H) |
17.7 x 25.2 x 37.4 in. |
---|---|
Filling volume | 48 ... 56 l |
Weight | 163.1 lbs |
Pump connections | M16x1 male |
Barbed fittings inner diameter | 8/12 mm |
Working temperature range | -38 ... +100 °C |
---|---|
Temperature stability | ±0.03 °C |
Setting the resolution of the temperature display | 0.1 °C |
Temperature display resolution | 0.01 ... 0.1 °C |
Ambient temperature | +5 ... +40 °C |
Heating capacity | 2 kW | |||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
||||||||||||||||||||||||
Pump capacity flow pressure | 17 l/min | |||||||||||||||||||||||
Circulation capacity pressure | 6.2 psi | |||||||||||||||||||||||
Viscosity max. | 50 cSt | |||||||||||||||||||||||
Power consumption | 14 A |
Bath tank | Stainless steel |
---|---|
Bath cover | integrated |
Usable bath opening (W x L / D) |
13.8 x 16.1 / 11.8 in. |
Cooling of compressor | 1-stage Air |
---|
Classification according to DIN12876-1 |
Classification III (FL) |
---|---|
IP Code according to IEC 60529 |
IP 21 |
Pump function | Pressure Pump |
Pump type | Immersion Pump |
Interfaces | USB |
---|---|
Temperature control | PID1 |
Absolute temperature calibration | 1 Point Calibration |
Temperature display | LED |
Temperature setting | Keypad |
9012708.03 | 230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A) |
---|---|
9012708.03.chn | 230V/50Hz (CN Plug) (R449A) |
9012708.04 | 230V/50Hz (UK Plug Type BS1363A) (R449A) |
9012708.05 | 230V/50Hz (CH Plug Type SEV 1011) (R449A) |
9012708.13 | 230V/60Hz (Nema N6-20 Plug) (R449A) |
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
Accessory for CORIO CD-1001F
Downloads for CORIO CD-1001F
We would also be happy to send you our printed catalogs and brochures by post.
Request brochureCORIO CD-1001F Benefits
Brilliant.
Very bright display makes it easy to read even from a distance.
Switch on. And off you go.
Intelligent operating concept. Ready for operation with just a few quick and easy steps.
Timer. Integrated.
CORIO circulators include an integrated timer function. When the set time has elapsed, a signal sounds and the device switches off. Setting range: 0 ... 999 minutes.
Everything at the front.
All operating controls and safety functions are accessed easily and comfortably from the front.
Related Products
Products with similar properties